Cleaning Instructions

67 | iASSIST 2-Pod Version Surgical Technique
IC Notice
This device complies with Industry Canada licence-exempt RSS standard(s). Operation is subject to the following two conditions:
(1) this device may not cause interference, and (2) this device must accept any interference, including interference that may cause
undesired operation of the device.
Cet appareil est conforme avec Industrie Canada exempts de licence standard RSS (s). Son fonctionnement est soumis aux deux
conditions suivantes: (1) cet appareil ne doit pas provoquer d’interférences et (2) cet appareil doit accepter toute interférence, y
compris celles pouvant causer un mauvais fonctionnement de l’appareil.
Health Canada RF Exposure Warning Statement
This device complies with Health Canada’s Safety Code. The installer of this device should ensure that RF radiation is not
emitted in excess of the Health Canada’s requirement. Information can be obtained at http://www.hc-sc.gc.ca/ewh-semt/pubs/
radiation/radio_guide-lignes_direct/index-eng.php
Cet appareil est conforme avec Santé Canada Code de sécurité 6. Le programme d ’installation de cet appareil doit s’ assurer
que les rayonnements RF n’est pas émis au-delà de I’exigence de Santé Canada. Les informations peuvent être obtenues: http://
www.hc-sc.gc.ca/ewh-semt/pubs/radiation/radio_guide-lignes_direct/index-fra.php
Guidance and Manufacturer’s Declaration
In order for the Tablet to attain its expected service life of three years, it must not be subjected to levels of electromagnetic distur-
bance higher than those specied in Tables 5.3, 5.4, 5.5 and 5.6. Failure to respect those levels may lead to a degradation of the
system’s basic safety and essential performances.
Table 5.3: Guidance and Manufacturer’s Declaration - ELECTROMAGNETIC EMISSIONS
The iASSIST V2 Tablet is intended for use in the electromagnetic environment specied below.
The customer or the user of the iASSIST Tablet should assure that it is used in such an environment.
Emissions test Compliance Electromagnetic environment – guidance
RF emissions
CISPR 11
Group 1 The iASSIST Tablet uses RF energy only for its internal function. Therefore, its RF emissions
are very low and are not likely to cause any interference in nearby electronic equipment.
RF emissions
CISPR 11
Class A The emissions characteristics of this equipment make it suitable for use in industrial areas
and hospitals (CISPR 11 class A). If it is used in a residential environment (for which CISPR
11 class B is normally required) this equipment might not oer adequate protection to ra-
dio-frequency communication services. The user might need to take mitigation measures,
such as relocating or re-orienting the equipment.
Caution: This equipment is not intended for use in residential environments and may not
provide adequate protection to radio reception in such environments.
Harmonic emissions
IEC 61000-3-2
Class D
Voltage uctuations/
icker emissions
IEC 61000-3-3
Complies
Table 5.4: Guidance and Manufacturer’s Declaration - ELECTROMAGNETIC IMMUNITY
The iASSIST V2 Tablet is intended for use in the electromagnetic environment specied below.
The customer or the user of the iASSIST Tablet should assure that it is used in such an environment.
IMMUNITY test
IEC 60601-1-2 test level
Compliance level Electromagnetic environment
– guidance
Electrostatic discharge (ESD)
IEC 61000-4-2
± 8 kV contact
± 2 kV, ± 4 kV, ± 8 kV, ± 15 kV air
± 8 kV contact
± 2 kV, ± 4 kV, ± 8 kV, ± 15 kV air
Floors should be wood, con-
crete or ceramic tile. If oors
are covered with synthetic
material, the relative humidity
should be at least 30 %.