Cleaning Instructions

65 | iASSIST 2-Pod Version Surgical Technique
5.6 Product Specications and Regulatory Notices
5.6.1 Essential Performances for the System
The System is composed of the iASSIST V2 kit and the iAssist Knee Pod Kit. The iAssist Pod contains inertial sensors
(accelerometers and gydroscopes) and instrument identication switches. The following performances have been identied as
essential to the system:
1. iASSIST V2 Tablet display.
2. Communication between the iASSIST V2 Tablet and the Pods contains accurate data. Complete loss of communication is not
considered an essential performance.
3. The accelerometer inclination measurements of the Pods on all axes are stable within ±1.0°.
4. The gyroscope rotation speed measurements of the Pods on all axes are stable within ±1.0°/sec.
5. Pod identication swtiches report accurate instrument connection.
6. Pod temperature changes are detected at ±1.0°C.
The essential performances 1, 3, 4,5 and 6 have been tested during electromagnetic immunity tests up to the levels shown
in the tables in this section. Loss or degradation of essential performance can result in navigation errors and/or instruments
misidentication.
5.6.2 iASSIST V2 Tablet Kit
Specications
- Weight: About 1.42 kg
- Dimensions: 335 mm x 220 mm x 43 mm (irregular)
- Power: 100-240 VAC, 47-63 Hz, 1.62-0.72 A
- Operating conditions: 15°C (5F) to 30°C (86°F), 20% to 90% relative humidity, 70 kPa to 106 kPa
- Transport conditions: -10°C (1F) to 49°C (120°F) at 10 to 90% non-condensing relative humidity
- Storage conditions: -10°C (1F) to 49°C (120°F) at 10 to 90% non-condensing relative humidity
- User devices: Touch screen, 2D barcode scanner
4. Remove the grey rubber keybpad.
5. Remove the circuit board.
6. Remove the foam and the battery