Cleaning Instructions

63 | iASSIST 2-Pod Version Surgical Technique
Status/Error LED Navigation LEDs Status Required Action
Pod Power-Up
Blinks rapidly (green) N/A The pods are trying to join the system
network for procedure
Wait for the blinking to slow down, indicating
that the network is up
Blinks slowly (green) N/A The network is up and ready for the
procedure
Proceed with surgery
Solid Red N/A Pod has failed self-test Open new pod kit
Calibration
Blinking green N/A The pod needs to be clipped to the
calibration jig for intraoperative
calibration or intraoperative calibration is
over and the pods can be unclipped from
the calibration jig to proceed with surgery
Follow instructions onscreen to clip or unclip
pods from the calibration jig
Solid green N/A The pods are adequately clipped to the
calibration and ready for the calibration
process
Proceed with surgery
N/A Solid green Indicates the expected calibration jig
position for the intraoperative calibration
Position the calibration jig so that the face
marked with that number is directed upward
and wait for the acquisition to be completed
Registration and Validation
N/A Blinking green Manipulations are needed to complete
the registration / validation acquisitions
Use the pattern of the LEDs or the indications
onscreen to perform necessary manipulation
N/A Steady green The leg is in the required orientation for
the next acquisition
Stabilize the leg in the actual position until
the registration sound is heard and another
movement is needed
Blinking green N/A Successful computation after a sequence
of acquisitions
No action needed. Continue with next step of
surgery.
Cut Guidance
N/A Solid red
Resection angle is out of the pod’s
range.
Use the gold (F/E -Slope) and green (V/V)
screws to adjust the resection angles.
N/A Solid red
Resection angle is not aligned (+/- 2
degrees) with target
Use the gold (F/E -Slope) and green (V/V)
screws to adjust the resection angles.
N/A Solid green
Resection angle is aligned (+/- 1 degree)
with target
Proceed with surgery
Anytime during procedure
Solid red N/A
Pod’s battery level is low
If the Pod battery icon displays two bars
instead of three, end of battery life within the
next 30 minutes.
N/A System or pod error Follow indications on the iASSIST V2 Tablet
display in order to resolve the error.
N/A End of battery life Consider opening a new iASSIST Knee Pod Kit
or switching to conventional instrumentation.
Blinking green N/A Everything is working as it should Proceed with surgery