Cleaning Instructions

61 | iASSIST 2-Pod Version Surgical Technique
5.5.1 iASSIT Tablet Kit (cont.)
Never insert foreign objects in the iASSIST Tablet
openings. This may cause electric shock or re, or
it may damage hardware components.
Make sure the computer is properly shutdown
before closing the iASSIST V2 Tablet Shipping Case.
If not, the system may overheat and be damaged.
The iASSIST Tablet must never be submerged in
water or exposed to high level of humidity.
Do not leave the iASSIST Tablet in an uncontrolled
environment where the storage temperature is
below -10°C (14°F) or above 49°C (120°F), 90%
relative humidity, non-condensing. This may
damage the equipment.
Make sure that no objects, including wires, are
left on the iASSIST Tablet screen when closing
the iASSIST V2 Tablet Shipping Case. Any objects
on the computer screen may damage it during
transportation or storage.
The iASSIST Tablet should not be used adjacent
to or stacked with other equipment. If adjacent
or stacked use is necessary, the tablet should
be observed to verify normal operation in the
conguration in which it will be used.
For safety reasons, the iASSIST Tablet must be
send to repair if the battery capacity is signicantly
reduced.
Because of risks related to lithium battery explosion,
the iASSIST Tablet should not be exposed to
extreme heat.
Do not use or store the iASSIST Tablet near a source
of heat or dust.
Do not use another power supply, or cables other
than the ones provided as this may negatively
aect EMC performances.
Make sure that the power cord provided with
the tablet is compatible with your local power
requirements.
Do not step on or place anything on the power
cord.
To avoid any damages to the power cables, make
sure they are not bent at sharp angles.
Never force a cable connection..
Touchscreen Calibration
No calibration is necessary for the iAssist Tablet V2
touch screen.
5.5.2 iASSIST Knee Pod Kit
The Pods should be stored at normal ambient
temperature (15°C to 30°C, 59°F to 86°F) and shipped
within temperatures from –18°C (-0.4°F) up to +49°C
(120°F), 80% relative humidity, non-condensing.
The Pods are packaged in a custom plastic tray
preventing activation prior to use.
Warning: Because of risks related to alkaline
battery leaking, the pods should not be exposed
to extreme heat such as an autoclave.
Care must be taken not to expose the pods to
strong impacts. Strong magnetic elds should not
approach the pods. This may erroneously trigger
switches aecting the surgical ow (accuracy of
sensors is not aected).
Functionality and accuracy of the pods may be
aected if exposed to X-ray radiation.
Be careful to not expose the pods to water
during the procedure. The pods can be wiped
with a cloth damped with sterile NaCL 0,9%. Any
excess water should be wiped o since it could
aect wireless communication.
Colorblind users may rely on the numbers displayed
next to navigation numbers. Also, the system
controller computer screen displays more details.
Work Around
In some rare condition, it is known that the pods
may lose network connection and may not be able
to reconnect to the system controller computer. In
this case, the user can re-scan the barcode, which
will reinitialize the RF session allowing lost pods to
reconnect the computer.