Cleaning Instructions

60 | iASSIST 2-Pod Version Surgical Technique
5.4 iASSIST V2 Tablet Kit
Before cleaning the iASSIST V2 Tablet kit, always
unplug and power o the Tablet from the AC power
outlet to prevent electrical shocks.
Clean the tablet screen only with a soft cloth dampened
with 60% or more isopropyl alcohol or 60% or more
ethyl alcohol. Wipe the screen and exterior with a soft,
damp cloth moistened only with water. Do not use
liquid or aerosol cleaners on the screen, as these will
discolor the nish and damage the screen.
Warning: Disinfecting computer equipment with
sprays is not recommended since the vapor can
enter the equipment which may cause electrical
short-circuits or corrosion. Never allow water or
other liquids to enter the tablet since this may
cause subsequent short-circuits or corrosion.
Hardware Product ID Qty Cleaning Instructions
iASSIST V2 Tablet 20-8011-070-19 1 See above
iASSIST V2 Tablet Power Supply 20-8011-070-20 1 See above
iASSIST V2 Tablet Shipping Case 20-8011-070-21 1 See above
iASSIST V2 Tablet Holder 20-8011-070-23 1 See above
5.5 General Equipment Information
Warning: If one of the following situations arises,
have the equipment veried by Zimmer CAS
customer support personnel:
a. The equipment is not functioning properly;
b. The iASSIST V2 Tablet or the instruments have
obvious/visible signs of breakage or fatigue;
c. The iASSIST V2 Tablet or the instruments have
been dropped and are damaged;
d. Liquid has entered the Tablet;
e. The power cord or plug is damaged; or
f. If the date and time function of the Tablet stops
functioning properly.
If an instrument marked with the below symbol is
dropped, have the instrument veried by Zimmer
CAS customer support personnel.
Note: When these instruments are dropped,
their precision may be impaired. The mechanical
deformity may be dicult to detect visually.
5.5.1 iASSIST V2 Tablet Kit
The iASSIST V2 Tablet has an internal battery that
allows a minimum of 30 minutes of autonomy. This
power backup is provided to prevent data loss if there
is a main power outage.
If the Tablet needs to be shipped with a carrier, place
it in the provided iASSIST V2 Tablet Shipping Case.
The iASSIST V2 Tablet Shipping Case is specically
designed to withstand carrier shipping conditions.
The expected life of the tablet under normal usage
conditions is three years.
Warning: Never open the casing of the iASSIST
Tablet. The tablet does not contain replaceable
parts. For safety reasons, only Zimmer CAS
customer support personnel should open the
equipment.
Protect the iASSIST Tablet from overheating. The
openings on the enclosure are for ventilation
purpose. Do not block the openings.