CLEANING AND CARE
CLEANING & CARE
SOLUTIONS
The following is a list of chemical cleaning solutions that have
been successfully tested on the Display Unit & Roll Stand:
• SANI-CLOTH HB
• SANI-CLOTH PLUS
• SUPER SANI-CLOTH
• Cavi Wipes
• Cloro-Wipe Towelette
• 70% Isopropyl Alcohol
• Alcohol Prep Pads
• Tuffie Wipes
DISPLAY UNIT
Use only a soft, lint-free cloth. Please note: abrasive cloths,
towels, paper towels or similar items should not be used as
these may cause damage to the Display Unit.
1. Disconnect the Display Unit from any external power sources.
2. Spray approved cleaning solutions onto soft lint-free cloth.
Do not spray cleaning solutions directly onto equipment or
at any openings.
3. Wipe surface until clean.
4. Cleaning of the Display Unit should be performed after
each use.
TRANSCEIVER
Wipe the Transceiver down with 70% isopropyl alcohol wipes
after each use.
ROLL STAND
(FOR LINKSTATION MINI ONLY)
The Roll Stand may be cleaned with most mild, non-abrasive
solutions commonly used in the hospital environment (e.g.
diluted bleach, ammonia, or isopropyl alcohol).
NOTE
The surface finish will be permanently damaged by strong
chemicals and solvents such as acetone and trichloro-
ethylene. Do not use steel wool or other abrasive material
to clean the Display Unit. Never submerge or allow liquids
to enter the Display Unit. Wipe any cleaning agents off the
Display Unit immediately using a water-dampened cloth.
Dry all Display Unit thoroughly after cleaning.