User's Guide

06/2019 LB-5200 Rev. 2
24
VERASENSE OrthoSensor.com
WARNING & DIALOG MESSAGES
VERASENSE Data Transfer Timeout.
Multiple VERASENSE found
SN
311717036
SN
412938089
Please select VERASENSE or ‘CANCEL’ to wait for
a different VERASENSE to become activated.
VERASENSE DATA TIMEOUT
This message will appear if an error has
occurred when handling measured data
that is being transferred from VERASENSE
for Exactech Equinoxe.
MULTIPLE VERASENSE FOUND
VERASENSE for Exactech Equinoxe selection
message when multiple VERASENSE for
Exactech Equinoxe sensors are active. Choose
desired VERASENSE for Exactech Equinoxe
by serial number (
SN
) listed on box.
VERASENSE Error. Unable to connect.
VERASENSE SHUTDOWN FAILED
This message will appear if an error
has occurred when trying to shutdown
VERASENSE for Exactech Equinoxe from
the Active Monitoring Screen. If
RETRY
is selected, the VERASENSE Software
Application for Shoulder will offer
additional shutdown attempts. If
EXIT
HOME SCREEN is selected then the
VERASENSE Software Application for
Shoulder will close and return to the
VERASENSE Software for Shoulder
Application Homepage.
VERASENSE ERROR - UNABLE
TO CONNECT
This message may appear when an error
has occurred during the initial connection
after the identification of VERASENSE
for Exactech Equinoxe and the VERASENSE
Software Application for Shoulder is
unable to properly obtain the VERASENSE
for Exactech Equinoxe EEPROM data.
Typical causes are due to loss of
VERASENSE for Exactech Equinoxe power
during activation or reception issues.
VERASENSE shutdown failed. Try again?
Error during VERASENSE Search. Select ‘OK’ to reset.
VERASENSE SEARCH - ERROR
This message appears when an error has
occurred while the VERASENSE Software
Application for Shoulder is searching for
active VERASENSE for Exactech Equinoxe.
Select OK and the VERASENSE Software
Application for Shoulder will reset and
re-establish the search for all active
VERASENSE for Exactech Equinoxe.