User's Guide
06/2019 LB-5200 Rev. 2
23
VERASENSE OrthoSensor.com
WARNING & DIALOG MESSAGES
VERASENSE Calibration Error.
VERASENSE overload detected.
VERASENSE CALIBRATION
ERROR
This message will appear if a VERASENSE
Software Application for Shoulder error
has occurred during the initial VERASENSE
for Exactech Equinoxe calibration while
on the Active Monitoring Screen.
VERASENSE OVERLOAD
DETECTED
VERASENSE for Exactech Equinoxe
overload condition in loads exceed
70lbf. Re-Zero should be performed.
VERASENSE error. Unable to connect.
VERASENSE ERROR -
UNABLE TO CONNECT
Error during the initial connection with
starting VERASENSE for Exactech Equinoxe.
Typically due to VERASENSE power loss
during connection.
VERASENSE Calibration Timeout.
VERASENSE CALIBRATION
TIMEOUT
This message will appear if a calibration
error occurs during Activation of VERASENSE
for Exactech Equinoxe.
• Select
RECALIBRATE to have the
VERASENSE Software Application for
Shoulder recalibrate VERASENSE for
Exactech Equinoxe prior to moving to
the Active Monitoring Screen.
• Select
SHUT DOWN VERASENSE to shut
down the current VERASENSE for
Exactech Equinoxe and to exit to the
VERASENSE Software Application for
Shoulder Home Screen.
Active VERASENSE found
(
SN
1234567890) Connect?
ACTIVE VERASENSE FOUND
This message appears when an active
VERASENSE for Exactech Equinoxe is
found. If the VERASENSE Software
Application is enabled with Channel
Switching and the VERASENSE Software
Application for Shoulder detects an
active VERASENSE for Exactech Equinoxe,
the following dialog box will appear
with the active VERASENSE for Exactech
Equinoxe serial number.
SERIAL NUMBER (
SN
) VERASENSE.
(The serial number is located on both the
VERASENSE for Exactech Equinoxe box as
well as the Tyvek Pouch that VERASENSE
for Exactech Equinoxe is in).