VERASENSE OrthoSensor.com
VERASENSE overload detected
Center of load and load values outside of the Green Zone are for reference only.
VERASENSE FOR EXACTECH EQUINOXE RE-ZERO LOAD
RE-ZEROING
If VERASENSE for Exactech Equinoxe experiences elevated loads,
a re-zero calibration may be required for accurate load readings.
A user can perform a re-zero as follows:
MENU ZERO BUTTON
Remove loaded VERASENSE for Exactech Equinoxe and then
press the Zero Button on the VERASENSE Software
Application for Shoulder.
Re-Zero Enabled
Re-Zero Completed
Center of load and load values outside of the Green Zone are for reference only.