User's Guide

06/2019 LB-5200 Rev. 2
10
VERASENSE OrthoSensor.com
VERASENSE FOR EXACTECH EQUINOXE ACTIVATION
ACTIVATE VERASENSE FOR
EXACTECH EQUINOXE
STEP 1
Select appropriate VERASENSE for Exactech Equinoxe size and remove
VERASENSE for Exactech Equinoxe and shims from outer box.
• DO NOT remove from sterile pouch.
STEP 2
Activate VERASENSE for Exactech Equinoxe by holding it stationary up
against the magnet.
NOTE: For the LinkStation MINI only, the magnet is mounted on the roll stand.
STEP 3
DO NOT REMOVE VERASENSE FOR EXACTECH EQUINOXE FROM THE
MAGNET UNTIL EACH OF THE FOLLOWING ACTIVATION STEPS OCCURS:
An LED light will illuminate in VERASENSE for Exactech Equinoxe
after approximately one (1) second. The light turns off after
approximately four (4) seconds.
The VERASENSE Software Application for Shoulder will recognize
VERASENSE for Exactech Equinoxe and initializes communication.
NOTE
To optimize battery life, do not activate VERASENSE for Exactech
Equinoxe until just before it is required in the surgical workflow.
VERASENSE for Exactech Equinoxe has a 40 minute battery life.
LINKSTATION MINI