Table 4 - Guidance and manufacturer’s declaration Electromagnetic Immunity and Emissions
The VERASENSE for Exactech Equinoxe and accessories have been tested and found to comply with the
electromagnetic compatibility (EMC) limits for medical devices to IEC 60601-1-2:2014.
Emissions/Immunity Test Compliance Level
RF Emissions
CISPR 11
Class A, Group 1
Electrostatic discharge (ESD)
IEC 61000-4-2
± 8 kV contact discharge
± 2, 4, 8 & 15kV air discharge
Radiated RF Immunity
IEC 61000-4-3
3 V/m
80 MHz to 2.7 GHz
80% at 1 kHz
Immunity to Proximity Fields from RF wireless
communication equipment
IEC 61000-4-3
28 V/m
Conducted Immunity
IEC 61000-4-6
0.15 - 80 MHz
3 Vrms & 6Vrms in ISM Band 1 kHz, AC Mains
Electrical fast transient/burst
IEC 61000-4-4
± 2 kV AC Mains
5kHz & 100 kHz
Magnetic Immunity
IEC 61000-4-8
30 A/m
Voltage dips, short interruptions and voltage
variations on power supply input lines
IEC 61000-4-11
0% U
T
.5 cycle
0% U
T
1 cycle
70% U
T
25 cycles
0% U
T
5 Sec