Use Instructions

OrthoSensor, Inc.
1855 Griffin Road
Suite A-310
Dania Beach, FL 33004-2200 USA
www.orthosensor.com
LB-5222 Rev. 1 (EN) 3
10-2019
For further information, contact the OrthoSensor Customer Service Center by phone at + 1 954-577-7770 or by e-mail at
customerservice@orthosensor.com or go to www.orthosensor.com.
Note: The 0mm or 2.5mm shim must be attached to the device prior to insertion into the humeral
component.
10. To remove the shim, or exchange for another size, simply unsnap the proximal lip of the attached shim and replace.
11. With the VERASENSE for Exactech Equinoxe and shim attached, physician should manually compress / apply load to
the device and verify the response on the User Interface prior to placing VERASENSE for Exactech Equinoxe into the
humeral tray.
12. Place the VERASENSE for Exactech Equinoxe within humeral tray.
13. Confirm that the VERASENSE for Exactech Equinoxe with shim is fully seated when placed in the humeral tray.
14. Flex the joint throughout its full range of motion to ensure appropriate response on the User Interface.
15. Proceed with rTSA process per physician / hospital protocol.
Note: If maximum allowable load of 70 lbf is reached, the VERASENSE for Exactech Equinoxe must be
removed from the shoulder joint and “re-zeroed” by pressing the Re-Zero button from the VERASENSE
Software Application for Shoulder, Re-Zero enabled will appear on the VERASENSE Software Application for
Shoulder, followed by Re-Zero Complete indicating that VERASENSE for Exactech Equinoxe has been reset to
zero.
16. Upon completion of the procedure, deactivate the VERASENSE Software Application for Shoulder by pressing the Exit
Button on the User Interface.
17. Dispose of the VERASENSE for Exactech Equinoxe per institutional guidelines for biohazardous medical waste.