ICS-2100 Ion Chromatography System
228 Doc. 065291-01 3/09
NOTE When the EGC-2 power supply is configured as an
auxiliary power supply, the EGC-2
Setpoint and
Actual Concentration fields display zeros.
B.8.3 Controlling the Eluent Generator Power
• To manually control the eluent generator power, touch the EGC Off
and
On buttons on the ELUENT GENERATOR page.
• To automatically control the eluent generator power simultaneously
with the pump flow, select
Auto turn on with pump.
NOTE The eluent generator power always turns off
when the pump is off, regardless of whether the
“Automatically turn on with pump” option is
selected.
NOTE When the EGC-2 power supply is configured as
an auxiliary power supply, the EGC-2 Off and
On buttons are not functional and the field
always displays
Off.
• To automatically control the auxiliary power supply simultaneously
with the pump flow, select
Auto turn on with pump for the EGC-2
and configure a TTL output for control of this function.
B.8.4 Monitoring the EluGen Cartridge Life
• The Expiration Date displayed on the ELUENT GENERATOR page is
2 years from the date of manufacture. Although you can continue
operation with the cartridge after the expiration date, performance
may be impaired until a new cartridge is installed.
NOTE When the EGC-2 power supply is configured as
an auxiliary power supply, the EGC-2
Expiration Date field is blank.
• Life Remaining (displayed when an EluGen cartridge is installed)
indicates the percentage of ions remaining in the cartridge. The ion
percentage is counted down in 1% increments. Warning messages are
displayed in the Chromeleon Audit Trail when the percentage falls