Home Theater System User Manual

4 Troubleshooting
Doc. 065291-01 3/09 97
Cell above or below temperature
Contact Dionex for assistance.
Damaged piston seal
Replace the piston seal (see Section 5.7
).
4.15 Vacuum Degas Assembly Does Not Run
Degas option not enabled in Chromeleon
1. Open the Chromeleon Server Configuration program. Right-click the
ICS-2100 device in the timebase and select Properties.
2. Select the Options tab (see Figure 4-3
).
.
Figure 4-3. ICS-2100 Server Configuration Properties: Options
3. Verify that the Degas check box is selected.