4.5 Pump Does Not Start
• No power (front Power LED indicator is not lit)
Check that the power cord is plugged in.
Check the main power fuses and replace, if needed (see Section 5.18
).
• No communication between ICS-2100 and Chromeleon
Verify that the USB cable is connected correctly. For connection and setup
information, refer to the ICS-2100 installation instructions.
• Pump is turned off
Turn on the pump from either the Chromeleon Control panel or the ICS-2100
touch screen
HOME page.
• Flow rate is set to 0
Set the flow rate from either the Chromeleon Control panel or the ICS-2100
touch screen
HOME page.
4.6 No Flow
• Pump waste or priming valve open (see Figure 4-2)
Close the valves by turning the knobs clockwise until fingertight. DO NOT
OVERTIGHTEN! Overtightening may damage the valve and the pump
head.