Home Theater System User Manual

ICS-2100 Ion Chromatography System
90 Doc. 065291-01 3/09
4.5 Pump Does Not Start
No power (front Power LED indicator is not lit)
Check that the power cord is plugged in.
Check the main power fuses and replace, if needed (see Section 5.18
).
No communication between ICS-2100 and Chromeleon
Verify that the USB cable is connected correctly. For connection and setup
information, refer to the ICS-2100 installation instructions.
Pump is turned off
Turn on the pump from either the Chromeleon Control panel or the ICS-2100
touch screen
HOME page.
Flow rate is set to 0
Set the flow rate from either the Chromeleon Control panel or the ICS-2100
touch screen
HOME page.
4.6 No Flow
Pump waste or priming valve open (see Figure 4-2)
Close the valves by turning the knobs clockwise until fingertight. DO NOT
OVERTIGHTEN! Overtightening may damage the valve and the pump
head.