User Manual
optris CS – E2012-10-B
4
Cautions
Avoid abrupt changes of the ambient temperature.
In case of problems or questions which may arise when you use the CS, please contact our service
department.
Read the manual carefully before the initial start-up. The producer reserves the right to change the herein
described specifications in case of technical advance of the product.
Factory Default Settings
The unit has the following presetting at time of delivery:
CS CSTC
Emissivity: 0,950 0,950
Transmission: 1,000 1,000
Average time: 0,3 s 0,3 s
Smart averaging: active active
Smart averaging hysteresis: 2 °C 2 °C
Ambient temperature source: internal (head) internal (head)
Status-LED function: Self diagnostic Self diagnostic
Input (IN/ OUT/ green): Communication input Communication input
Output (OUT/ yellow): mV output inactive
Temperature range: 0...350 °C -40…1030°C
Output voltage: 0...3,5 V according to t/c type K
Thermocouple output: - active
Vcc adjust: inactive inactive
Signal processing: Hold mode: off Hold mode: off
Calibration: Gain 1,000/ Offset 0,0 Gain 1,000/ Offset 0,0
Failsafe: inactive inactive
The default settings can be
changed with the optional USB kit
(USB adapter cable + software).
If the unit is supplied together with
the USB kit the output is already
preset to digital communication
(bidirectional).