User Manual
optris CS – E2012-10-B
19
Electrical Installation
Please use a stabilized power supply unit with an output voltage in the range of 5–30 VDC which can supply
100 mA. The residual ripple should be max. 200 mV.
Analog device (mV output on OUT pin)
Analog device (thermocouple type K output on OUT t/c K pins / only at model CSTCLT15)
The output impedance must be ≥ 10kΩ.
The output impedance must be ≥ 20 Ω.