User Manual

Emissivity Setting
After opening of the sensor backplane [ Cable Connections] both of the emissivity switches are accessible.
For an emissivity setting of 1,00 please turn both switches to 0. Values below 0,10 are not adjustable. For all
other switch positions the following applies: 0, S1
S2.
Therefore the adjustment range is 0,10...1,09.
Example: ε = 0,84 S1=8
optris CSlaser – E2010-07-B
23
S
2=4
If you use the software (optional) please consider that the emissivity
switches can be activated/ deactivated in the software menu Device/
Device setup. At time of delivery the switches are active.
The emissivity set in the software interacts as a factor to the emissivity set
on the unit. Thus the adjustment range increases to 0,100...1,199.
Example: ε
Software
= 0,952/ ε
Sensor
= 0,82 (S1=8/ S2=2)
Therefore the effective emissivity is: 0,781.