[Cartridge troubleshooting]
For problems occurred while using the LOAA cartridges, refer to the possible solutions below. If cartridge leaks, refer to
chapter 0 for proper decontamination procedures.
Problem
Possible cause of error
Solution
Cartridge pouch's
individual package is not
sealed.
Cartridge contamination and moisture
buildup is possible.
Discard the cartridge and test the sample using a
new cartridge.
The sample to be loaded
into the cartridge does
not start to spread out
over the well area after
starting POSTMAN
operation.
There may be bubbles in the cartridge
inlet after sample injection.
Discard the cartridge and test the sample using a
new cartridge.
Operation failure Cartridge does not function properly.
Test with the new cartridge.
If operation fails repeatedly, Please, contact the
technical support team for assistance.