Use Instructions

Instructions For Use
Page 24 / 47
IFU-3M001-EN-00 (2021.03.05)
LOAA(Lab on An Array) Real-Time PCR Instrument
5.4.2. Validating well structures inside the Dr. PCR Cartridge
Figure 15. Verifying Cartridge Screen
Figure 16. Analyzing Cartridge Screen
Cartridge analysis HOME>New Test>Sample ID>Next>Next
1) When you press Next button on the screen, the loader automatically moves back in the equipment and the door is
closed.
2) It automatically checks the cartridge status (completed within about 30 seconds).
Caution:
If there is no problem in the cartridge to be used, it will proceed to the next step to start test, but if there is any
error message occurred, please, refer to section 7 for error code specific to cartridge or device problems for
further action.
5.4.3. Start the test
When cartridge verification and analysis is done successfully, test will start automatically.
Figure 17. Test screen
Test Progress Start Test>Test process>Temperature
1) Press the Temp. button on the Temperature and Process Cycle screen of the test to see temperature and cycle
progress during the test.