Use Instructions

Instructions For Use
Page 20 / 47
IFU-3M001-EN-00 (2021.03.05)
LOAA(Lab on An Array) Real-Time PCR Instrument
Figure 10. Operating LOAA Sample Loader
7) The sample is automatically loaded so that uniform volume is injected into the wells in the cartridge.
8) When the sample loading is finished, the LOAA Sample Loader (POSTMAN) automatically stops running and the
button will blink in random colors. (it takes about 60 seconds)
9) Press the cartridge guide toward the center to release the cartridge from POSTMAN (Figure 11).
Figure 11. Separate Dr. PCR Cartridge
10) Hold the cartridge with both hands and lock it in place by pressing the top and bottom cases in various directions
as shown in Figure 12.
Figure 12. Dr. PCR Cartridge Lock direction
5.2.4. Starting to operate LOAA Analyzer
LOAA software includes:
1) Guides the user to place the cartridge in the LOAA Analyzer
2) Checking the problems of the cartridge installed and,
3) Depending on cartridge types, importing cartridge information from the memory inside the cartridge or input the
information from user. To start test, touch New Test on the home screen.