4. Technical Data
Specification Item
Specification
Performance
Performance Item
Specification
sample processing 1 test at a time per LOAA, 1 sample / test
Light source 460nm LED
Heating ramp rates (max.)
10℃/sec from 60℃ to 95℃
Cooling ramp rates (max.)
5℃/sec from 95℃ to 60℃
Temperature accuracy
Within ±0.5℃
Maximum power consumption Less than 90 W
Operating conditions
• Temperature: 15℃~30℃
• Relative humidity: 20 to 80% (noncondensing)
• Ambient pressure: 700hPa to 1,060hPa
• Indoor use only
Storage conditions
• Temperature: 2℃ to 50℃ (35.6℉ to 122℉) in manufacturer’s package
• Relative humidity: 20% to 80% humidity
• Ambient pressure: 700hPa to 1,060hPa
Transport conditions
• Temperature: -10℃ to 50℃ (14℉ to 122℉) in manufacturer’s package
• Relative humidity: 10% to 85% humidity
• Ambient pressure: 700hPa to 1,060hPa
Power requirements • 100-240V~ 50~60Hz 1.3A input, 24 Vdc. output, 3.75A, 90W
Dimensions and weight
• 242 x 253 x 206 mm (H x D x W)
• Total weight: 4.1 Kg
- Analyzer: 3.2 Kg (With adaptor: 3.7 Kg)
- LOAA Sample Loader (Model name: POSTMAN) 0.34 Kg
EMC requirements
LOAA Analyzer system complies with the emission and immunity requirements of IEC 61326-
1: 2012 (Class A).
CPU Cortex-A53 Octa core 1.4 GHz
Storage and Memory
• 2GB DDR3
• 8GB eMMC
• 1TB HDD
Interface and peripherals
• One Ethernet network interface
• 2 USB ports for storage (Located on the lower back)
• Wi-Fi antenna: 5Ghz/2.4Ghz
Display
5.5” TFT LCD with touch
• 68.04x120.96mm active area
• 1920x1080 FHD resolution
• multi-touch support
Operating system Android 7.1 (Nougat)
Software version 3.0.0
Overvoltage category II