LOAA(Lab on An Array) Real-Time PCR Instrument
2. Installation of LOAA Analyzer for Cartridge System
2.1. Installation requirements
Choose a location for LOAA Analyzer system with a clean, well-ventilated flat table where you can place LOAA Analyzer.
• Keep space of at least 15 cm between the product and the wall.
• Power specifications: 100~240 Vac, 50/60Hz 1.3A (Output: 24 Vdc, 3.75A) use a rated adaptor.
LOAA Analyzer complies with the emission and immunity requirements of IEC 61326-1. It is recommended that you evaluate
the electromagnetic environment before operating the device.
Caution!
Do not position the LOAA Analyzer and LOAA Sample Loader in such a way that it is difficult to disconnect
the power adapter and cord set from the devices.
2.2. LOAA Analyzer System Components
Each system is provided with devices, Instructions For Use, DC adapter and cord. The contents of each box are shows in
Figure 3.
Figure 3.LOAA Analyzer Components
Note The LOAA Sample Loader (Model name: POSTMAN) is delivered separately in its own box. USB-A to USB-C power
cable is provided and any USB power adaptor with output power rating of 5V, 1A or more can be used as a power supply
for LOAA Sample Loader (Model name: POSTMAN).