User Manual
OPR-3301 SS08xxx
17
12.5. Electrical Specifications
Specifications mentioned below include combination of the scanner and CDR-3301.
- Withstand Voltage: AC 1500 V/ per 60 seconds, 10mA or less
- Insulation Resistance: DC 500 V, 2 MΩ or higher
- Current Leakage: 250 μA or less / AC 250 V 60 Hz
- Power Line Noise Immunity: ± 1kV and higher
- Static Electricity Noise Immunity:
No destruction found: ± 15kV (air or direct discharge)
No malfunction found: ± 8 kV (air or direct discharge),
±6kV (contact, direct or indirect discharge)
Note: Testing method complies with IEC-61000-4-2. 150pF, 330Ω
12.6. Drop Test (Without Individual Packaging)
No malfunction occurred after the following drop test.
Drop Test: Drop the scanner from a height of 150 cm onto a concrete floor (three times in each of 5
angles as shown in the figure.
150cm
Figure 10: Drop test
12.7. Drop Test (With Individual Packaging)
No malfunction occurred after the following drop test.
Drop Test: Drop an individually packaged scanner from a height of 150 cm onto a concrete floor
once on its 1 corner, 3 edges, and 6 sides (10 total drop tests)
12.8. Vibration Strength
No malfunction occurred after the following vibration test.
Vibration test: Increase the frequency of the vibration from 10 Hz to 100 Hz with accelerated
velocity 19.6m/s2 (2G) for 30 minutes (60 minutes for one cycle) in non-operating state. Repeat
this routine in each X, Y, Z directions.
Concrete floor