Operating Instructions | Revision 06/2014 | Print | Item No.
30 V variant item no. 11011 115 V variant item no. 11021 230 V variant item no. 11012 115 V variant item no. 11022 These operating instructions apply to the footcare systems PODOLOG ECO (item no. 11011) and PODOLOG ONE (item no. 11012), referred to in the descriptions that follow as the PODOLOG ECO / ONE.
Dear Customer, Thank you for choosing a PODOLOG ECO or PODOLOG ONE. This high-quality product from RUCK will provide optimum support for your day-to-day work, whether based in a practice or on the move. Backed by our name and our two-year guarantee, you can be assured of many years of enjoyable and problem-free operation. Please take some time to read through these operating instructions carefully before setting up the system.
TABLE OF CONTENTS 1. Device description .................................................7 1.1 1.1.1 1.1.2 1.1.3 Explanation of the characters and icons used ....................... 9 On the device and transport packaging 9 In the operating instructions 9 Display warnings 10 1.2 Scope of delivery ............................................................... 10 1.3 Packaging .......................................................................... 10 1.4 Recycling ................................
4. Set-up.................................................................17 4.1 5. Technical safety ................................................................ 17 Operation ............................................................20 5.1 5.1.1 5.1.2 5.1.3 5.1.4 5.1.5 Preparations for operation .................................................
.3 Cleaning the motor unit ...................................................... 38 6.3.1 Cleaning the motor unit 39 6.4 7. Cleaning and disinfecting the controller .............................. 41 Accessories and spare parts.................................. 42 7.1 Tools ................................................................................. 42 7.2 Dust filter bags .................................................................. 42 7.3 Foot controller .................................
1.
01 02 03 04 Ventilation slots Foot switch connection Device lock on base Power cable If the power cable is damaged, the device must be disconnected from the mains and may not be used any more. A damaged power cable may only be replaced by an authorised specialist.
1.1 EXPLANATION OF THE CHARACTERS AND ICONS USED 1.1.
1.1.3 DISPLAY WARNINGS "HOLDER BLOCKED!" - If the holder is blocked, the motor speed and suction turbine are switched off after about 5 seconds. This warning disappears after about 10 seconds, after which the holder speed can be reset. "EXTRACTION OVERHEATING" - If the permissible air temperature in the suction turbine housing is exceeded, the suction turbine and holder are switched off. Once the temperature has fallen by around 10 - 15 °C, this warning disappears and the holder speed can be reset. 1.
Please keep the packaging (box, plastic bag, foam) to use for returning the device in case of any maintenance work. 1.4 RECYCLING Old equipment must be disposed of as electronic waste and should not be treated as domestic waste. Full dust bags can be disposed of with general rubbish. Please make sure you follow any local guidelines in this respect. Special country-specific requirements should also be taken into account.
2. INTENDED USAGE The PODOLOG ECO and PODOLOG ONE footcare systems operate rotating tools such as nail file drills and grinders for the removal of hard skin, calluses, nails, etc. from feet and hands. The device is intended for use in the following fields of activity: Footcare, nail design, cosmetics, wellness and related professions. Any other usage is undertaken at your own risk and could be dangerous. Improper usage of the device can lead to personal and material damage.
catalogue from HELLMUT RUCK GMBH, in the section "Rotary instruments". − Operators should not come into contact with the foot controller connection and the client at the same time. − Protective low voltage must be maintained. − Operators must wear eye, mouth and nose protectors and gloves while using the equipment. − During treatment, the operator must ensure that hair or loose items such as towels, tamponades, etc. cannot get onto the area of the rotary tools.
If any germ-intensive work has been carried out, disinfected and sterilised tools must be used during the next work process, to treat any unaffected areas of skin or nail. 3.1.1 ALLERGY RISKS In the case of clients with intolerance to components made of medicinal steel, there exists a possibility, albeit a small one, of an allergic reaction.
3.1.3 CLIENT INJURY CAUSED BY HEAT, OUTPUT OF ENERGY AND MATTER, HEAT AS A RESULT OF FRICTION ON THE SKIN OR NAILS Heat development through friction on the skin is a problem, particularly when working at higher rev. counts with large-surface nail file drills or tools. In cases such as these, the working and contact surfaces should not be too small. You shouldn't do deep, point-focused work but you should make wider, sideways cutting movements across the skin.
The efficient extraction of the PODOLOG ECO / ONE greatly improves visibility and reduces the risk of accidental deeper injuries. 3.2 OPERATOR RISKS The extraction of potentially pathogenic dust significantly reduces the risks caused by transferable germs (micro-organisms). Risks to the operator through inhalation are kept to a minimum. For reasons of hygiene we recommend the wearing of face/mouth protection and gloves. 3.2.
3.2.3 INJURY CAUSED BY HAIR, JEWELLERY OR CLOTHING COMING INTO CONTACT WITH ROTARY TOOLS For reasons of working safety, it is vital that long hair is tied back closely to the head or covered with a cap or hat. For reasons of hygiene, the wearing of jewellery and watches should be avoided. Working clothes should be selected so that no fringing or loose material can get caught up in the tool. 4. SET-UP 4.
− Do not store the PODOLOG ECO / ONE directly next to or with other devices. Should this be necessary, please make sure that all intended functions of the PODOLOG ECO / ONE work properly. The use of other equipment than that prescribed may result in increased emissions or reduced immunity to interference. Risk of strangulation from holder hose and power cable. Keep children away from the device. Risk of swallowing in the case of required accessories (small parts), e.g. nail file drills.
The instructions regarding operation and transport of the device must be adhered to. Please ensure that you have easy access to the mains plug at all times so that the device can be unplugged from the mains without difficulty. The power plug represents the separator to the mains power supply.
5.
01 Standby button Press this button to activate/deactivate the standby mode 02 Extraction power setting Increase [+] to decrease [-] the extraction power. The power level is shown in the display. 03 Info menu [ i ] Through this menu, you can for example, activate the holder cleaning process or check when the next filter change or service is due. Press and hold the button for 5 seconds to change base settings, e.g. language.
− Close the drawer by sliding it in. − Switch the device on using the standby button.
You should always work with a dust filter bag in place and securely connected. The filter bag drawer must always be closed when the unit is in operation. The PODOLOG ECO / ONE switches off if the drawer is opened. 5.1.1 INSERTING A ROTARY TOOL Insert the rotary tool (grinder/nail file drill) that you need for your work in the opening of the holder. The tool only has to be inserted (quick-action chuck).
The first time it is switched on, the PODOLOG ECO / ONE is set to your language and your area of activity. Additionally, you can activate the service reminder function so that you are reminded in time when the next service is due. Please refer to Chapter 6, MAINTENANCE AND CLEANING. You are guided through the Start menu, step-by-step. At the end, you confirm your inputs with the [M1] button.
THANK YOU! [-] BACK [M1] FINISHED If you want to modify any of the Start menu settings in the future, press and hold the [ i ] button for 5 seconds.
5.1.3 SWITCHING THE HOLDER ON Set the required rotational speed using the speed control knob. The displayed number x 1000 indicates the rpm value. Grip the holder and operate the switch at the rear end of the holder. Switch Switch The motor runs at the set speed and you can start your work. The speed setting may not exceed the permitted maximum speed of your tool. 5.1.4 SWITCHING THE HOLDER OFF While the motor is running, operate the switch at the rear end of the holder.
5.1.5 SWITCHING OFF THE PODOLOG ECO / ONE Press the standby button. The PODOLOG ECO / ONE is now switched off and goes into stand-by mode. Please ensure that you have easy access to the mains plug at all times so that the device can be unplugged from the mains without difficulty. The power plug represents the separator to the mains power supply. 5.2 SETTINGS 5.2.1 HOLDER MOTOR SPEED The rotational speed that corresponds to the tool being used is set using the rotational speed knob.
5.2.2 RIGHTHAND / LEFTHAND DIRECTION OF ROTATION You can set your tool to righthand or lefthand rotation depending on which tool you are using and on whether you are right-handed or lefthanded. Righthand rotation The arrow shows the direction of rotation. Arrow to the right – Righthand rotation Arrow to the left – Lefthand rotation 5.2.3 MODIFYING THE ROTATION DIRECTION The rotation direction can be modified while the machine is running. Press the arrow button.
5.2.4 SELECTING THE EXTRACTION LEVEL You can set the extraction power level to meet the current working requirements. Repeated pressing of the [ - ] button decreases the extraction power level. Repeated pressing of the [ + ] button increases the extraction power level. Pressing and holding the corresponding button controls the extraction power level in the desired direction. There are five extraction power levels. 5.2.
SAVED With a short press on [M1], [M2] or [M3], the corresponding stored value is called. You can change the settings at any time while the machine is in operation. The speed setting may not exceed the permitted maximum speed of your tool.
5.3 − − − TOOL CHANGING Switch off the motor at the holder. Pull the tool out of the holder. Insert a different tool into the holder opening. Tools should only be changed when the holder motor has been switched off. The tool shaft must be fully inserted! 5.4 ERGONOMIC HOLDER GRIP 5.4.1 FOUNTAIN PEN GRIP Fine work can be done ergonomically and accurately using the fountain pen grip. 5.4.
5.5 INFO MENU The Info menu offers you the option of calling up the following information and functions. Scroll through the individual menu items with the [ + ] and [ - ] buttons. 5.5.1 FILTER CHANGE If you want to change the fine dust filter, you can do this using the Info menu and inform the device at the same time that a new filter has been inserted. You will find an extensive description under Point 6.2.1 CHANGING THE FILTER BAG on page 36.
5.5.2 DISPLAY NEXT SERVICE DATE This practical display lets you know when the next service is due for your PODOLOG ECO / ONE: DEVICE TYPE: Service after: due PODOLOG ECO PODOLOG ONE 2 years or 900 operating hours 1 year or 500 operating hours INFO MENU NEXT RECOMMENDED SERVICE: 01/20.. OR AFTER …OPERATING HOURS [+/-] SCROLL [i] MAIN MENU . 5.5.3 ACTIVATING THE SERVICE REMINDER If you activate the Service reminder function, your PODOLOG ECO / ONE will let you know when the next service is due.
INFO MENU DEACTIVATE SERVICE REMINDER? PLEASE PRESS [M1] [+/-] SCROLL 6. MAINTENANCE AND CLEANING 6.1 MAINTENANCE [i] MAIN MENU To a great extent, the PODOLOG ECO / ONE is maintenance-free. The most important aspect of maintenance is the timely changing of the filter. It guarantees a correct suction function. You should clean the outside of the device and disinfect it if required after each treatment and remove dust particles from the holder at least once a day.
HELLMUT RUCK GmbH Daimlerstraße 23 75305 Neuenbürg Germany Tel. +49 (0) 7082 944 20 Fax +49 (0) 7082 944 22 22 kontakt@hellmut-ruck.
6.2 FILTER CHANGE A filter change needs to be carried out at the latest after operating the holder for 30 hours. The device gives you a reminder through the filter change icon in the display, that the dust filter is due to be changed. Select: = NOW = LATER If you select LATER, the device displays a reminder every 5 hours until the filter change has been carried out. 6.2.
FILTER CHANGE PLEASE CHANGE FILTER [M1] NOW [M2] LATER Start the change process with the [M1] button: FILTER HAS A NEW FILTER BEEN INSTALLED? [M1] YES [M2] CANCEL Open the dust filter bag drawer by pulling on the hose connector. Hold the filter drawer in such a way that the hose connector still points upwards. Now remove the full bag from the hose sleeve. Close it using the plug supplied for this purpose. Insert a new bag and locate the opening over the hose sleeve. Close the dust bag drawer.
The dust filter bag should not be emptied and reused. Should you use the device for the treatment of artificial nails (nail design, nail prosthesis) please take into consideration that the time between filter changes shifts due to the high degree of erosion. A weekly check of the level is necessary to ensure smooth operation.
The specifications and instructions from the manufacturer referring to cleaning and disinfectant materials and those referring to the corresponding devices must be followed. 6.3.1 CLEANING THE MOTOR UNIT Do not use disinfecting sprays. Dampness could penetrate and lead to bearing damage. We recommend RUCK alcohol-free disinfectant cloths (item no 2942801) for gentle cleaning and disinfection. The holder can be disassembled for cleaning.
Motor unit Motor unit Never immerse the device in water or other fluids as this creates a risk of an electric shock.
Do not use disinfecting sprays. Dampness could penetrate and lead to damage to the electronics. We recommend RUCK alcohol-free disinfectant cloths (item no 2942801) for gentle cleaning and disinfection. 6.4 CLEANING AND DISINFECTING THE CONTROLLER The controller can only be cleaned externally using a special plastics cleaner (e.g. RUCK plastics cleaner, item no. 10335) and a surface disinfecting agent. Never immerse the device in water or other fluids as this creates a risk of an electric shock.
7. ACCESSORIES AND SPARE PARTS Please note! The accessories listed here are only to be used with the PODOLOG ECO / ONE . Using them with other medical-electrical devices can lead to increased emissions or reduced immunity to interference for the device in question. 7.1 TOOLS You can use all rotary instruments (nail file drill, grinder, polisher, etc) in accordance with DIN EN ISO 1797-1 with normed, 2.35mm ∅ shafts. You will find a wide range in our catalogue. 7.
8. TECHNICAL DATA PODOLOG ECO 230 V version 115 V version Controller dimensions (W/D/H) in mm 250/235/120 250/235/120 Weight in g approx. 2800 approx. 2800 Holder dimensions Length in mm 127 127 Weight in g approx. 110 approx. 110 Voltage Controller 230 V, 50 Hz AC voltage 115 V, 60 Hz AC voltage Power consumption max. 200 VA max. 200 VA Micromotor voltage 24 V 24 V Constant load 40 W 40 W Idling speed (rpm) 25 000 25 000 Volume flow at max. power* approx. 190 I/min.
PODOLOG ONE 230 V version 115 V version Controller dimensions (W/D/H) in mm 250/235/120 250/235/120 Weight in g approx. 3300 approx. 3300 Holder dimensions Length in mm 122 122 Weight in g approx. 110 approx. 110 Voltage Controller 230 V, 50 Hz AC voltage 115 V, 60 Hz AC voltage Power consumption max. 350 VA max. 350 VA Micromotor voltage 24 V 24 V Constant load 40 W 40 W Idling speed (rpm) 25 000 25 000 Volume flow at max. power* approx. 155 I/min. approx. 155 I/min.
9. GUARANTEE HELLMUT RUCK GmbH (RUCK) gives a two year guarantee on the PODOLOG ECO / ONE. Normal wear and tear, in particular to bearings, plug-connections, suction hoses and clamps, is excluded. The guarantee runs from the date of purchase when the product was first bought. The date of purchase has to be confirmed by the sales receipt. A claim can only be made in combination with the device logbook.
10. DISPOSAL ENVIRONMENTAL HAZARD! Incorrect disposal endangers our environment. >>> Please use available return and collection systems for disposing of the PODOLOG ECO / ONE at the end of its life-cycle! Address HELLMUT RUCK GmbH Daimlerstraße 23 75305 Neuenbürg Germany 46 Tel. +49 (0) 7082 944 20 Fax +49 (0) 7082 944 22 22 kontakt@hellmut-ruck.