User Manual

©2010–2012OnsetComputerCorporation.Allrightsreserved.Onset,HOBO,andHOBOwarearetrademarksorregisteredtrademarksofOnsetComputerCorporationforitsdatalogger
productsandconfiguration/interfacesoftware.Allothertrademarksarethepropertyoftheirrespectivecompanies.
13416CMANU1400x
U14DataLogger
(ModelU14001shown)
HOBO
®
U14 Data Logger User Manual
TheUfamilyofdataloggersoffersreliabilityandconvenientmonitoringforapplicationsthat
requirehigheraccuracy,betterresolution,morememory,orUSBconnectivityforfastdata
readouttoyourcomputer.
TheHOBOU14loggerrecordsanddisplaystemperatureandhumidityconditionsin
manufacturing,processing,andstorageenvironmentswherereliablemonitoring
and
documentationofspecifictemp/RHconditionsarecritical.
TheHOBOU14isavailableintwoversions:U14001withinternalsensorsandtheU14002
withexternalsensors.TheHOBOU14requiresHOBOware
®
ProorHOBOwareLite,Version2.5
orlater.
Features
64knonvolatilememorythatretainsdataevenifthebatteriesfail
MemoryindicatoronLCDshowshowmuchdatastoragespaceremains
BatteryLevelindicatorshowsremainingusefulbatterylife
OperatesforoneyearonthreealkalineAAAbatteriesintypicalconditions
Builtinrelay
forusewithRemoteAlarmorAutomaticDialer
DirectUSBinterfaceforlaunchinganddatareadoutusingHOBOware
Sample(logging)intervalcanbesetfromonesecondto18hours
SupportsOnset’sDataAssistants(availablewithHOBOwaresoftware)
CanbeusedwiththeOnsetUShuttle
(UDT1)
Internalsensormodel(U14001)hasinternaltemperatureandrelativehumiditysensorswith12bitresolution
Externalsensormodel(U14002)supportsthe12bitTemperature/10bitRHSensor(STHBM00x)andthe12bit
TemperatureSensor(STMBM00x)
Accessories
OutdoorCase(CASE4x)
RemoteAlarm(ARA)
AutomaticDialer(AVD45)
LocationRequirements
TheHOBOU14isdesignedforanoncondensing,noncorrosiveenvironment.Keeptheloggerdry.Someexternalsensorscanbe
placedinharsherenvironmentsthanthelogger.Seetheexternalsensorspecificationsinthesensorusermanual.
Specifications
Environmental
Operatingtemperaturerange‐20°Cto+50°C,(4°Fto+122°F),0to100%RHnoncondensing
Operatingrelativehumidityrange 0to95%,noncondensing,noncorrosiveenvironment
LoggerCapabilities
Capacity TempandRH:21,500measurements;TempOnly:43,000measurements
Softwareselectablesamplingintervals 1secondupto18hours,recordingtimesupto1year
Start Programmablestarttime/date
Recordingmodes Stopwhenfull
LaunchModes Immediate,Delayed
Memory NonvolatileEEPROMmemoryretainsdataevenifbatteriesfail