Operation Manual
109
Detailed information about your system
9
Precision
Within Run Precision (300 Venous Blood Samples Tested
per Glucose Level)
Data generated using the OneTouchVerio Flex® Meter.
Target
Glucose
(mg/dL)
Mean
Glucose
(mg/dL)
Standard
Deviation
(mg/dL)
Coecient
of Variation
(%)
40 36.5 0.83 2.26
90 88.5 1.72 1.95
130 127.9 2.50 1.95
200 199.3 3.94 1.98
350 344.6 6.18 1.79
Results show that the greatest variability observed
between test strips when tested with blood is 0.83mg/dL
SD or less at glucose levels less than 75mg/dL or 1.98% CV
or less at glucose levels at 75mg/dL or above.