Operation Manual

107
Detailed information about your system
9
Fingertip Results for Glucose
Concentrations <75mg/dL
Percent (and number) of meter results that match the
laboratory test
Within ±5mg/dL Within ±10mg/dL Within ±15mg/dL
76.2% (16/21) 95.2% (20/21) 95.2% (20/21)
Fingertip Results for Glucose
Concentrations 75mg/dL
Percent (and number) of meter results that match the
laboratory test
Within ±5% Within ±10% Within ±15% Within ±20%
58.3%
(88/151)
88.7%
(134/151)
98.0%
(148/151)
99.3%
(150/151)