Operation Manual

53
    MESSAGE WHAT IT MEANS
Error message that
indicates that the
blood or control
solution sample
was applied before
Apply Blood,
Alternate Site, or
Control Solution
appeared on the
display.
1) You may have
high glucose and
have tested in
an environment
near the low end
of the system’s
operating
temperature
range (43–111°F/
6–44°C).
ACTION
Repeat the test
with a new test
strip. Apply the
blood or control
solution sample
only after Apply
Blood”, Alternate
Site, or Control
Solution” appears
on the display. Refer
to pages 33–35
for information on
sample application.
1) If you tested in a
cool environment,
repeat the test
in a warmer
environment with
a new test strip.
If you were
testing in a
normal or warm
environment,
repeat the test
with a new test
strip (see pages
7–8 for test strip
information).
WA RNING
See Owner’s Booklet
Error 4
Strip problem
Please repeat test
WA RNING
See Owner’s Booklet
Error 3
Meter not ready
Please repeat test