Operation Manual

89
Troubleshooting your system
8
What it means
Meter is too cold (below 6°C)
to perform a test.
What to do
Move the meter and test strips to a warmer area. Insert a
new test strip when the meter and test strips are within
the operating range (6-44°C). If you do not get another
LO.t message, you can proceed with testing.
If this message continues to appear, contact Customer
Service. Contact OneTouch® Customer Care on
0800121200 (UK) or 1800535 676 (Ireland).