Operation Manual

81
6
Caring for your system
Disinfecting your meter, lancing device
and cap
The meter, lancing device and cap should be disinfected
periodically. Clean your meter, lancing device and cap
prior to disinfecting. For disinfecting, obtain regular
household bleach (containing a minimum of 5.5% sodium
hypochlorite as the active ingredient)*. Prepare a solution
of 1 part household bleach and 9 parts water.
*Follow manufacturer’s instruction for handling and
storage of bleach.
1. Hold the meter with the test strip port pointed down
Use a soft cloth dampened
with this solution to wipe
the outside of the meter
and lancing device until the
surface is damp. Be sure to
squeeze out any excess liquid
before you wipe the meter.
2. After wiping, cover the surface you are disinfecting
with the soft cloth dampened with the bleach solution
for 1 minute
Wash hands thoroughly with
soap and water after handling
the meter, lancing device and
cap.
If you see signs of wear, please contact Customer Service.
Contact OneTouch® Customer Care on 0800121200 (UK)
or 1800535 676 (Ireland).