Yellow= 5mm Red=7mm margin Ultra® Plus Flex Blood Glucose Monitoring System Owner's Booklet 05983201A_OT_UPF_OB_mmolL_Inside_R2.
Yellow= 5mm Red=7mm margin Thanks for choosing OneTouch®! The OneTouch Ultra Plus Flex™ Blood Glucose Monitoring System is one of the latest product innovations from OneTouch®. Your OneTouch Ultra Plus Flex™ Meter is designed to connect (sync) with a variety of devices running software applications that let you review and graph your results, and help identify patterns. Meter results are sent to the compatible devices either through BLUETOOTH® SMART (wireless) or USB cable connection.
Yellow= 5mm Red=7mm margin Compatible Wireless Devices Visit www.LifeScan.co.uk for information on which wireless devices are compatible with your OneTouch Ultra Plus Flex™ Meter, and where/how to download the software application on your compatible wireless device. Meter symbols and icons Meter Power Low Battery Battery Empty Control Solution Syncing BLUETOOTH® SMART Feature On History Mode (Past results) Apply Sample Range Indicator Arrow Setting Mode 05983201A_OT_UPF_OB_mmolL_Inside_R2.
Yellow= 5mm Red=7mm margin Other symbols and icons Cautions and Warnings: Refer to the Owner's Booklet and inserts that came with your system for safety-related information. Direct current Consult Instructions for Use Manufacturer Lot Number Serial Number Storage Temperature Limits In Vitro Diagnostic Device Do Not Re-use Sterilized by irradiation Not for general waste Use By Date Contains sufficient for n tests Underwriters Laboratories certification 05983201A_OT_UPF_OB_mmolL_Inside_R2.
Yellow= 5mm Red=7mm margin Before you begin Before using this product to test your blood glucose, carefully read this Owner's Booklet, and the inserts that come with the OneTouch Ultra® Plus Test Strips, OneTouch Ultra® Plus Control Solutions and the OneTouch® Delica® Lancing Device. IMPORTANT SAFETY INSTRUCTIONS: • This meter and lancing device are for single patient use only.
Yellow= 5mm Red=7mm margin Intended use The OneTouch Ultra Plus Flex™ Blood Glucose Monitoring System is intended to be used for the quantitative measurement of glucose (sugar) in fresh capillary whole blood samples drawn from the fingertip. The system is intended to be used by a single patient and should not be shared.
Yellow= 5mm Red=7mm margin Test principle Glucose in the blood sample mixes with the enzyme FAD-GDH (see page 100) in the test strip and a small electric current is produced. The strength of this current calculates your blood glucose level. It then displays the blood glucose result and stores it in the meter memory. Use only OneTouch Ultra® Plus Control Solutions and Test Strips with the OneTouch Ultra Plus Flex™ Meter.
Yellow= 5mm Red=7mm margin Visit www.OneTouch.com for information on which wireless devices are compatible with your OneTouch Verio Flex™ Meter, and where/how to download the software application on your compatible wireless device. When using the OneTouch Verio Flex™ System, we suggest you pair your OneTouch Verio Flex™ Meter with a compatible wireless device and track your results. See page 29 for pairing instructions.
Yellow= 5mm Red=7mm margin These guidelines help ensure that your meter will not affect the operation of other nearby electronic devices. Additionally, other electronic devices should not affect the use of your meter. WARNING: The BLUETOOTH® SMART feature on your meter sends test results to your compatible wireless device. To prevent other people's results from being sent to your compatible wireless device, Do Not let anyone else use your meter to test their blood glucose.
Yellow= 5mm Red=7mm margin Table of Contents 1 Getting to know your system....................................12 2 Setting up your system............................................. 22 Setting up your meter...................................................... 22 Connecting to a compatible wireless device ................ 27 Turning the meter off....................................................... 33 3 Taking a test...............................................................
Yellow= 5mm Red=7mm margin 5 Editing Your Settings................................................ 74 Editing time and date...................................................... 74 Editing your range limits................................................. 75 6 Caring for your system............................................. 78 7 Battery........................................................................ 82 8 Troubleshooting your system..................................
Yellow= 5mm Red=7mm margin 1 Getting to know your system Your OneTouch Ultra Plus Flex™ Blood Glucose Monitoring System Included with your kit: Lancing device OneTouch Ultra Plus Flex™ Meter (CR2032 lithium coin cell battery included) 05983201A_OT_UPF_OB_mmolL_Inside_R2.
Yellow= 5mm Red=7mm margin Getting to know your system 1 NOTE: If any item is missing or defective in your kit, contact Customer Service. Call OneTouch® Customer Care on 0800 121 200 (UK) or 1800 535 676 (Ireland) during the hours of 8:30am-6pm Monday-Friday, 9am-1pm Saturday. Or visit www.LifeScan.co.uk. NOTE: If another type of lancing device was included, see the separate instructions for that lancing device. 05983201A_OT_UPF_OB_mmolL_Inside_R2.
Yellow= 5mm Red=7mm margin 1 Getting to know your system Available separately: Items pictured below are required, but may not be included in your kit: They are sold separately. Refer to your meter carton for a list of included items. OneTouch Ultra® OneTouch Ultra® Plus Mid Control Plus High Control Solution* Solution* (vial with blue cap) (vial with red cap) OneTouch Ultra® Plus Test Strips* *OneTouch Ultra® Plus Control Solution and Test Strips are available separately.
Yellow= 5mm Red=7mm margin Getting to know your system 1 WARNING: Keep the meter and testing supplies away from young children. Small items such as the battery door, batteries, test strips, lancets, protective covers on the lancets, and control solution vial cap are choking hazards. Do Not ingest or swallow any items. 05983201A_OT_UPF_OB_mmolL_Inside_R2.
Yellow= 5mm Red=7mm margin 1 Getting to know your system Getting to know your OneTouch Ultra Plus Flex™ Blood Glucose Monitoring System 05983201A_OT_UPF_OB_mmolL_Inside_R2.
Yellow= 5mm Red=7mm margin Getting to know your system 1 05983201A_OT_UPF_OB_mmolL_Inside_R2.
Yellow= 5mm Red=7mm margin 1 Getting to know your system Getting to know your OneTouch Ultra® Plus Test Strip Test strip 05983201A_OT_UPF_OB_mmolL_Inside_R2.
Yellow= 5mm Red=7mm margin Getting to know your system 1 The Range Indicator feature The OneTouch Ultra Plus Flex™ Meter automatically lets you know if your current result is below, above or within your range limits. It does this by displaying your current result with a Range Indicator Arrow, pointing to a corresponding Range Indicator Color Bar below the meter display. Use the Range Indicator Arrow and Color Bar together to interpret your results. 05983201A_OT_UPF_OB_mmolL_Inside_R2.
Yellow= 5mm Red=7mm margin 1 Getting to know your system 3 Possible Range Indicator Displays A Range Indicator Arrow will appear just below your result after each test depending upon how you set your low and high range limits in the meter. Things you should know before using the Range Indicator feature: • The meter comes with pre-set range limits. The pre-set low range limit is 3.9 mmol/L and the pre-set high range limit is 10.0 mmol/L. You can change these limits as needed to meet your needs.
Yellow= 5mm Red=7mm margin Getting to know your system 1 This page left blank intentionally. 05983201A_OT_UPF_OB_mmolL_Inside_R2.
Yellow= 5mm Red=7mm margin 2 Setting up your system Setting up your meter Turn your meter on To turn your meter on, press and hold until the start-up test screen appears. Once the device is on, release . You can also turn the meter on by inserting a test strip. Every time you turn your meter on, a start-up screen will appear for a few seconds. All segments of the display should appear briefly, indicating your meter is working properly. If the meter does not power on, check the battery.
Yellow= 5mm Red=7mm margin Setting up your system 2 First time setup until the startTo turn your meter on, press and hold up screen appears. Once the device is on, release . The meter will now automatically prompt you to set the time and date. The SET icon will appear on the screen to indicate that the meter has entered setup mode. Setting the time Hour will flash • If the hour displayed is correct, press to confirm. • If the hour displayed is not correct, press or to change the hour and press to confirm.
Yellow= 5mm Red=7mm margin 2 Setting up your system Setting the date After completing the time setup, the meter will automatically move to date setup. Year will flash • If the year displayed to is correct, press confirm. • If the year displayed is not correct, press or to change the year and press to confirm. Month will flash • If the month displayed to is correct, press confirm. • If the month displayed is not correct, press or to change the month and press to confirm.
Yellow= 5mm Red=7mm margin Setting up your system 2 Day will flash • If the day displayed is to correct, press confirm. • If the day displayed is not correct, press or to change the day and press to confirm. You are now ready to take a test. See the section Testing your blood glucose in Chapter 3. NOTE: After completing first time setup, a screen with three dashes will appear.
Yellow= 5mm Red=7mm margin 2 Setting up your system Adjusting the time and date settings after first time setup You can adjust the meter's time and date settings after to turn the meter on, first time setup. Press and hold then press and hold and at the same time. The SET screen will appear. See page 74. After adjusting the settings, your meter will exit settings mode and your last result screen will appear. 05983201A_OT_UPF_OB_mmolL_Inside_R2.
Yellow= 5mm Red=7mm margin Setting up your system 2 Connecting to a compatible wireless device Turning the BLUETOOTH® SMART feature on or off In order to connect your meter with your compatible wireless device, the BLUETOOTH® SMART feature will need to be turned on. The symbol will appear on the meter screen when the BLUETOOTH® SMART feature is on. When the symbol is not present on the screen the BLUETOOTH® SMART feature is off. • To turn the BLUETOOTH® SMART feature on press and at the same time.
Yellow= 5mm Red=7mm margin 2 Setting up your system Pairing Overview Pairing allows your OneTouch Ultra Plus Flex™ Meter to communicate with compatible wireless devices. The devices must be within 8 meters of each other to pair and sync. Download the OneTouch Reveal® Mobile App from the appropriate app store before pairing your meter and compatible wireless device. NOTE: Some diabetes management apps, including the OneTouch Reveal® Mobile App, may not be available in your country. Visit www.LifeScan.co.
Yellow= 5mm Red=7mm margin Setting up your system 2 Pairing Instructions 1. Start by turning your meter on using the button 2. The BLUETOOTH® SMART feature is turned on by and together pressing The symbol will appear to indicate that the BLUETOOTH® SMART feature is on. 3. Open the OneTouch Reveal® Mobile App and follow instructions to pair meter with your compatible wireless device 05983201A_OT_UPF_OB_mmolL_Inside_R2.
Yellow= 5mm Red=7mm margin 2 Setting up your system 4. Look for "OneTouch" and the last 4 characters of the meter serial number on the compatible wireless device display to correctly identify your meter 05983201A_OT_UPF_OB_mmolL_Inside_R2.
Yellow= 5mm Red=7mm margin Setting up your system 2 5. When prompted by the OneTouch Reveal® Mobile App, the meter will display a six digit PIN number Enter the PIN number into your wireless compatible device using the keypad on your compatible wireless device. CAUTION: Make sure the PIN you enter on your compatible device matches the PIN on your meter display.
Yellow= 5mm Red=7mm margin 2 Setting up your system Syncing to send results wirelessly to the OneTouch Reveal® Mobile App After pairing the meter with your compatible wireless device, you are ready to send results to the OneTouch Reveal® Mobile App. 1. Open the OneTouch Reveal® Mobile App on your compatible wireless device to turn the meter on and make sure 2.
Yellow= 5mm Red=7mm margin Setting up your system 2 Turning the meter off There are three ways to turn your meter off: for several seconds until the meter • Press and hold turns off. • Remove the test strip. • Your meter will turn off by itself if left alone for two minutes. NOTE: Following a glucose test, the meter will still be available for BLUETOOTH® SMART connection for up to 4 hours. See page 69 for more details. 05983201A_OT_UPF_OB_mmolL_Inside_R2.
Yellow= 5mm Red=7mm margin 3 Taking a test Testing your blood glucose Preparing for a blood glucose test NOTE: Many people find it helpful to practice testing with control solution before testing with blood for the first time. See page 58. Have these things ready when you test: OneTouch Ultra Plus Flex™ Meter OneTouch Ultra® Plus Test Strips Lancing device Sterile lancets NOTE: • Use only OneTouch Ultra® Plus Test Strips.
Yellow= 5mm Red=7mm margin Taking a test 3 • Do Not test if there is condensation (water build-up) on your meter. Move your meter and test strips to a cool, dry spot and wait for the meter surface to dry before testing. • Do Not open the test strip vial until you are ready to remove a test strip and perform a test. Use the test strip immediately after removing it from the vial. • Tightly close the cap on the vial immediately after use to avoid contamination and damage.
Yellow= 5mm Red=7mm margin 3 Taking a test • Comparing your blood glucose test results taken with this meter to your results taken from a different meter is not recommended. Results may differ between meters and are not a useful measure of whether your meter is working properly. To check your meter accuracy, you should periodically compare your meter results to those obtained from a lab. See page 96 for more information.
Yellow= 5mm Red=7mm margin Taking a test 3 • If you cannot test due to a problem with your testing supplies, contact your healthcare professional. Failure to test could delay treatment decisions and lead to a serious medical condition. • The test strip vial contains drying agents that are harmful if inhaled or swallowed and may cause skin or eye irritation.
Yellow= 5mm Red=7mm margin 3 Taking a test Getting to know your OneTouch® Delica® Lancing Device 05983201A_OT_UPF_OB_mmolL_Inside_R2.
Yellow= 5mm Red=7mm margin Taking a test 3 NOTE: • The OneTouch® Delica® Lancing Device uses ONLY OneTouch® Delica® Lancets. • If another type of lancing device was included, see the separate instructions for that lancing device. • The OneTouch Ultra Plus Flex™ Blood Glucose Monitoring System has not been evaluated for Alternate Site Testing (AST). Use only fingertips when testing with the system.
Yellow= 5mm Red=7mm margin 3 Taking a test Lancing precautions CAUTION: To reduce the chance of infection and disease spread by blood: • Make sure to wash the sample site with soap and warm water, rinse and dry before sampling. • The lancing device is intended for a single user. Never share a lancet or lancing device with anyone. • Always use a new, sterile lancet each time you test. • Always keep your meter and lancing device clean (See page 78).
Yellow= 5mm Red=7mm margin Taking a test 3 Preparing the lancing device 1. Remove the lancing device cap Remove the cap by turning it counterclockwise and then pulling it straight off of the device. 05983201A_OT_UPF_OB_mmolL_Inside_R2.
Yellow= 5mm Red=7mm margin 3 Taking a test 2. Insert a sterile lancet into the lancing device Align the lancet as shown here, so that the lancet fits into the lancet holder. Push the lancet into the device until it snaps into place and is fully seated in the holder. Twist the protective cover one full turn until it separates from the lancet. Save the protective cover for lancet removal and disposal. See page 55. 05983201A_OT_UPF_OB_mmolL_Inside_R2.
Yellow= 5mm Red=7mm margin Taking a test 3 3. Replace the lancing device cap Place the cap back onto the device; turn clockwise to secure the cap. Do Not overtighten. 05983201A_OT_UPF_OB_mmolL_Inside_R2.
Yellow= 5mm Red=7mm margin 3 Taking a test 4. Adjust the depth setting The lancing device has seven puncture depth settings, numbered 1 through 7. Smaller numbers are for a shallower puncture and the larger numbers are for a deeper puncture. Shallower punctures work for children and most adults. Deeper punctures work well for people with thick or callused skin. Turn the depth wheel to choose the setting. NOTE: A shallower fingertip puncture may be less painful.
Yellow= 5mm Red=7mm margin Taking a test 3 Preparing the meter 1. Insert a test strip to turn the meter on Insert a test strip into the test strip port with the gold side of the test strip and the two silver prongs facing you. NOTE: No separate step to code the meter is required. The flashing blood drop icon ( ) appears on the display. You can now apply your blood sample to the test strip. 05983201A_OT_UPF_OB_mmolL_Inside_R2.
Yellow= 5mm Red=7mm margin 3 Taking a test Getting a blood sample from the fingertip Choose a different puncture site each time you test. Repeated punctures in the same spot may cause soreness and calluses. Before testing, wash your hands thoroughly with warm, soapy water. Rinse and dry completely. 1. Puncture your finger Hold the lancing device firmly against the side of your finger. Press the release button. Remove the lancing device from your finger. 05983201A_OT_UPF_OB_mmolL_Inside_R2.
Yellow= 5mm Red=7mm margin Taking a test 3 p 2. Get a round drop of blood ss Gently squeeze and/or massage your fingertip until a round drop of blood forms on your fingertip. m, NOTE: If the blood smears or runs, Do Not use that sample. Dry the area and gently squeeze another drop of blood or puncture a new site. 05983201A_OT_UPF_OB_mmolL_Inside_R2.
Yellow= 5mm Red=7mm margin 3 Taking a test Applying blood and reading results Applying the sample 1. Apply the sample to the test strip You can apply blood to either side of the test strip. Apply your sample to the opening of the channel. Be sure to apply your sample immediately after you get a drop of blood. Holding the meter at a slight angle, guide the channel to the blood drop. 05983201A_OT_UPF_OB_mmolL_Inside_R2.
Yellow= 5mm Red=7mm margin Taking a test 3 When it touches your sample, the test strip wicks blood into the channel. a 2. Wait for the channel to fill completely The blood drop will be drawn into the narrow channel. The channel should fill completely. The channel turns red and the meter will count down from 5 to 1. Blood should not be applied on the top of the test strip or to the top edge of the test strip. 05983201A_OT_UPF_OB_mmolL_Inside_R2.
Yellow= 5mm Red=7mm margin 3 Taking a test NOTE: • Do Not smear or scrape the sample with the test strip. • Do Not press the test strip too firmly against the puncture site or the channel may be blocked from filling properly. • Do Not apply more blood to the test strip after you have moved the drop of blood away. • Do Not move the test strip in the meter during a test or you may get an error message or the meter may turn off.
Yellow= 5mm Red=7mm margin Taking a test 3 Viewing your result Your result appears on the display, along with the unit of measure, and the date and time of the test. After your glucose result appears, the meter will also display a Range Indicator Arrow below your glucose result to indicate if your result is below, above or within your range limits (see page 19). The arrow will point to the appropriate Range Indicator Color Bar on the meter as a visual reminder.
Yellow= 5mm Red=7mm margin 3 Taking a test Interpreting unexpected results Refer to the following cautions when your results are higher or lower than what you expect. CAUTION: Low results If your result is below 3.9 mmol/L or is shown as LO (meaning the result is less than 1.1 mmol/L), it may mean hypoglycemia (low blood glucose). This may require immediate treatment according to your healthcare professional's recommendations.
Yellow= 5mm Red=7mm margin Taking a test 3 CAUTION: High results • If your result is above 10.0 mmol/L, it may mean hyperglycemia (high blood glucose) and you should consider re-testing. Talk to your healthcare professional if you are concerned about hyperglycemia. • HI is displayed when your result is over 33.3 mmol/L. You may have severe hyperglycemia (very high blood glucose). Retest your blood glucose level. If the result is HI again, this indicates a severe problem with your blood glucose control.
Yellow= 5mm Red=7mm margin 3 Taking a test CAUTION: Repeated unexpected results • If you continue to get unexpected results, check your system with control solution. • If you are experiencing symptoms that are not consistent with your results and you have followed all instructions in this Owner's Booklet, call your healthcare professional. Never ignore symptoms or make significant changes to your diabetes management program without speaking to your healthcare professional.
Yellow= 5mm Red=7mm margin Taking a test 3 Removing the used lancet NOTE: This lancing device has an ejection feature, so you do not have to pull out the used lancet. 1. Remove the lancing device cap Remove the cap by turning it counterclockwise and then pulling it straight off of the device. 2. Cover the exposed lancet tip Before removing the lancet, place the lancet protective cover on a hard surface then push the lancet tip into the cupped side of the cover. 05983201A_OT_UPF_OB_mmolL_Inside_R2.
Yellow= 5mm Red=7mm margin 3 Taking a test 3. Eject the lancet Slide the ejection control forward until the lancet comes out of the lancing device. Return the ejection control to its back position. If the lancet fails to eject properly, cock the device again and then slide the ejection control forward until the lancet comes out. 05983201A_OT_UPF_OB_mmolL_Inside_R2.
Yellow= 5mm Red=7mm margin Taking a test 3 4. Replace the lancing device cap Place the cap back onto the device; turn clockwise to secure the cap. Do Not overtighten. It is important to use a new lancet each time you obtain a blood sample. Do Not leave a lancet in the lancing device. This will help prevent infection and sore fingertips. 05983201A_OT_UPF_OB_mmolL_Inside_R2.
Yellow= 5mm Red=7mm margin 3 Taking a test Disposing of the used lancet and test strip Discard the used lancet carefully after each use to avoid unintended lancet stick injuries. Used lancets and test strips may be considered biohazardous waste in your area. Be sure to follow your healthcare professional's recommendations or local regulations for proper disposal. Wash hands thoroughly with soap and water after handling the meter, test strips, lancing device and cap.
Yellow= 5mm Red=7mm margin Taking a test 3 • Tightly close the cap on the control solution vial immediately after use to avoid contamination or damage. See page 34 for more information on testing. CAUTION: • Do Not swallow or ingest control solution. • Do Not apply control solution to the skin, eyes, ears or nose as it may cause irritation. • Do Not use control solution after the expiration date (printed on the vial label) or the discard date, whichever comes first, or your results may be inaccurate.
Yellow= 5mm Red=7mm margin 3 Taking a test When to do a control solution test • When you open a new vial of test strips. • If you suspect that the meter or test strips are not working properly. • If you have had repeated unexpected blood glucose readings. • If you drop or damage the meter. Preparing your meter for a control solution test 1. Insert a test strip to turn the meter on 05983201A_OT_UPF_OB_mmolL_Inside_R2.
Yellow= 5mm Red=7mm margin Taking a test 3 2. Wait for the flashing blood drop icon ( ) to appear on the display or until the 3. Press control solution icon ( ) appears on the display on Preparing the control solution 1. Before removing the cap, shake the vial gently 2. Remove the vial cap and place it on a flat surface with the top of the cap pointing up 3. Squeeze the vial to discard the first drop 05983201A_OT_UPF_OB_mmolL_Inside_R2.
Yellow= 5mm Red=7mm margin 3 Taking a test 4. Wipe both the tip of the control solution vial and the top of the cap with a clean, damp tissue or cloth 5. Squeeze a drop into the small well on the top of the cap or onto another clean, non-absorbent surface 05983201A_OT_UPF_OB_mmolL_Inside_R2.
Yellow= 5mm Red=7mm margin Taking a test 3 Applying the control solution 1. Hold the meter so that the side edge of the test strip is at a slight angle to the drop of control solution 2. Touch the channel on the side of the test strip to the control solution 3. Wait for the channel to fill completely 05983201A_OT_UPF_OB_mmolL_Inside_R2.
Yellow= 5mm Red=7mm margin 3 Taking a test Viewing your control solution result After the control solution is applied, the meter will count down until the test is complete. Your result is displayed along with the date, time, unit of measure, and (for control solution) and stored in the meter. Control solution results can be seen when reviewing past results on the meter. CAUTION: Make sure you press or until the control appears before you begin a control solution solution icon test.
Yellow= 5mm Red=7mm margin Taking a test 3 Checking if the result is in range Compare the result displayed on the meter to the range printed on your OneTouch Ultra® Plus Control Solution vial. If your control solution result falls outside the expected range, repeat the test with a new test strip. Example range OneTouch Ultra® Plus Mid Control Solution Control Range 5.7-7.7 mmol/L OneTouch Ultra® Plus High Control Solution Control Range 16.5-22.
Yellow= 5mm Red=7mm margin 3 Taking a test Causes of out-of-range results Out-of-range results may be due to: • Not following the instructions for performing a control solution test. • Control solution is contaminated, expired, or past its discard date. • Test strip or test strip vial is damaged, expired, or past its discard date. • Meter, test strips and/or control solution were not all at the same temperature when the control solution test was performed. • A problem with the meter.
Yellow= 5mm Red=7mm margin Taking a test 3 Cleaning the control solution cap Clean the top of the control solution cap with a clean, damp tissue or cloth. 05983201A_OT_UPF_OB_mmolL_Inside_R2.
Yellow= 5mm Red=7mm margin 4 Reviewing past results Reviewing past results on your meter Your meter stores your most recent 500 blood glucose and control solution test results and displays them in the order the tests were taken. The ( ) symbol will appear on your screen when in History Mode. 1. When the meter is off, press and hold History Mode on to turn The ( ) symbol indicates you are viewing your past results.
Yellow= 5mm Red=7mm margin Reviewing past results 4 Using the meter without syncing to an app The meter can be used without a compatible wireless device or the app. You can still test your blood glucose and review up to 500 results on the meter. Sending your results to the app If the BLUETOOTH® SMART feature on the meter is turned on, indicated by the BLUETOOTH® SMART symbol ( ), the meter will automatically send the latest result to any paired wireless compatible device.
Yellow= 5mm Red=7mm margin 4 Reviewing past results To ensure that glucose test results are successfully sent to the app, turn on the BLUETOOTH® SMART feature and check the following: • The compatible wireless device and meter are both turned on, and the app is running. • The meter is correctly paired with your compatible wireless device. • The Bluetooth feature on both devices is running (indicated by ) and the devices are within 8 meters of each other.
Yellow= 5mm Red=7mm margin Reviewing past results 4 If you are still unable to send results to the compatible wireless device, please call Customer Service. Contact OneTouch® Customer Care on 0800 121 200 (UK) or 1800 535 676 (Ireland). NOTE: Inserting a test strip during the transmission will cancel the transfer of all results. The symbol appears on the screen and you can proceed with testing. 05983201A_OT_UPF_OB_mmolL_Inside_R2.
Yellow= 5mm Red=7mm margin 4 Reviewing past results Downloading results to a computer Your meter can work with diabetes management software, which provides a visual way to track key factors that affect your blood sugar. To learn more about diabetes management tools available to you, contact Customer Service. Call OneTouch® Customer Care on 0800 121 200 (UK) or 1800 535 676 (Ireland) during the hours of 8:30am-6pm Monday-Friday, 9am-1pm Saturday. Or visit www.LifeScan.co.uk.
Yellow= 5mm Red=7mm margin Reviewing past results 4 are, 0 it ). Once the command to start the download is sent from the computer to the meter, the meter display will show the flashing Sync symbol ( ) indicating that the meter is in communication mode. Do Not insert a test strip while the meter is connected to a computer. If you are unable to download your results to a computer, please call Customer Service. Contact OneTouch® Customer Care on 0800 121 200 (UK) or 1800 535 676 (Ireland).
Yellow= 5mm Red=7mm margin 5 Editing Your Settings Editing time and date You can adjust the meter's time and date settings after first time setup. Press and hold to turn the meter on, then press and hold and at the same time. The SET screen will appear, and the hour will flash. For instructions on adjusting the time and date, see page 23. After adjusting the settings, you will exit settings mode and your last glucose result screen will appear.
Yellow= 5mm Red=7mm margin Editing Your Settings 5 Editing your range limits Your meter uses low and high range limits to tell you when your result is below, above or within your set range. The meter comes with pre-set range limits that can be changed. The pre-set low range limit is 3.9 mmol/L and the pre-set high range limit is 10.0 mmol/L. To edit the pre-set range limits press and hold and at the same time.
Yellow= 5mm Red=7mm margin 5 Editing Your Settings 1. Review the pre-set low range limit displayed • To accept the pre-set low range limit, press . • To edit the pre-set low range limit, press or to change the value between 3.3 mmol/L-6.1 mmol/L, and then press . 2. Review the pre-set high range limit displayed • To accept the pre-set high range limit, press . • To edit the pre-set high range limit, press or to change the value between 5.0 mmol/L-16.7 mmol/L, and then press .
Yellow= 5mm Red=7mm margin Editing Your Settings 5 NOTE: If you change your range limits, this will only affect whether future test results are displayed as below, above or within your range limits. Changing your range limits does not affect how past results are displayed. NOTE: You will not be able to perform a glucose test until you finish editing the range limits. NOTE: You can use the OneTouch Reveal® Mobile App on your compatible wireless device to change the range limits stored in your meter.
Yellow= 5mm Red=7mm margin 6 Caring for your system Storing your system Store your meter, test strips, control solution and other items in your carrying case. Keep in a cool, dry place between 5°C and 30°C. Do Not refrigerate. Keep all items away from direct sunlight and heat. Cleaning and disinfection Cleaning and disinfection are different and both should be performed. Cleaning is part of your normal care and maintenance and should be performed prior to disinfection, but cleaning does not kill germs.
Yellow= 5mm Red=7mm margin Caring for your system 6 • Do Not use alcohol or any other solvent. • Do Not allow liquids, dirt, dust, blood or control solution to enter the test strip port or the data port. (See page 16.) • Do Not spray cleaning solution on the meter or immerse it in any liquid. 1.
Yellow= 5mm Red=7mm margin 6 Caring for your system 2. Wipe dry with a clean, soft cloth Disinfecting your meter, lancing device and cap The meter, lancing device and cap should be disinfected periodically. Clean your meter, lancing device and cap prior to disinfecting. For disinfecting, obtain regular household bleach (containing a minimum of 5.5% sodium hypochlorite as the active ingredient)*. Prepare a solution of 1 part household bleach and 9 parts water.
ce d m on Yellow= 5mm Red=7mm margin Caring for your system 6 1. Hold the meter with the test strip port pointed down Use a soft cloth dampened with this solution to wipe the outside of the meter and lancing device until the surface is damp. Be sure to squeeze out any excess liquid before you wipe the meter. 2.
Yellow= 5mm Red=7mm margin 7 Battery Replacing the battery Your OneTouch Ultra Plus Flex™ Meter uses one CR2032 lithium coin cell battery. IMPORTANT: Use only one CR2032 lithium coin cell battery with your meter. Do Not use rechargeable batteries. Use of an incorrect battery type may result in your meter providing fewer tests than normal. If the meter does not turn on, you may need to replace the battery. See below for instructions.
tery e Yellow= 5mm Red=7mm margin Battery 7 1. Remove the old battery Start with the meter turned off. Remove the battery cover by sliding it downward. the Pull up on the battery ribbon to lift the battery out of the compartment. Do Not remove the battery while the meter is connected to a computer. 05983201A_OT_UPF_OB_mmolL_Inside_R2.
Yellow= 5mm Red=7mm margin 7 Battery 2. Insert the new battery Insert one CR2032 lithium coin cell battery on top of the battery ribbon, with the plus (+) side up. If the meter does not power on after you have replaced the meter battery, check that the battery is correctly installed. If the meter still does not power on, contact Customer Service. Contact OneTouch® Customer Care on 0800 121 200 (UK) or 1800 535 676 (Ireland). 3.
Yellow= 5mm Red=7mm margin Battery 7 4. Check your meter settings Removing the meter battery will not affect your stored results. However, you will need to check your date and time settings. 5. Dispose of battery Dispose of the battery according to your local environmental regulations. 05983201A_OT_UPF_OB_mmolL_Inside_R2.
Yellow= 5mm Red=7mm margin 8 Troubleshooting your system Error and other messages The OneTouch Ultra Plus Flex™ Meter displays messages when there are problems with the test strip, with the meter or when your glucose levels are above 33.3 mmol/L or below 1.1 mmol/L. Improper use may cause an inaccurate result without producing an error message. NOTE: If the meter is on but does not operate (locks-up), contact Customer Service.
Yellow= 5mm Red=7mm margin Troubleshooting your system 8 What it means You may have a very high blood glucose level (severe hyperglycemia), over 33.3 mmol/L. NOTE: When your glucose result is above 33.3 mmol/L, both the HI and the Range Indicator Arrow will flash on the meter screen. What to do Re-test your blood glucose level. If the result is HI again, obtain and follow instructions from your healthcare professional right away. 05983201A_OT_UPF_OB_mmolL_Inside_R2.
Yellow= 5mm Red=7mm margin 8 Troubleshooting your system What it means Meter is too hot (above 44°C) to perform a test. What to do Move the meter and test strips to a cooler area. Insert a new test strip when the meter and test strips are within the operating range (6-44°C). If you do not get another HI .t message, you can proceed with testing. If this message continues to appear, contact Customer Service. Contact OneTouch® Customer Care on 0800 121 200 (UK) or 1800 535 676 (Ireland).
Yellow= 5mm Red=7mm margin Troubleshooting your system 8 What it means Meter is too cold (below 6°C) to perform a test. What to do Move the meter and test strips to a warmer area. Insert a new test strip when the meter and test strips are within the operating range (6-44°C). If you do not get another LO.t message, you can proceed with testing. If this message continues to appear, contact Customer Service. Contact OneTouch® Customer Care on 0800 121 200 (UK) or 1800 535 676 (Ireland).
Yellow= 5mm Red=7mm margin 8 Troubleshooting your system Error Screens If there is a problem with your meter, there are six possible error screens that may appear. Along with an error number, there is also an error code in the upper left corner of your meter screen. If you cannot resolve the error with your meter, contact Customer Service. Contact OneTouch® Customer Care on 0800 121 200 (UK) or 1800 535 676 (Ireland). They will refer to the error number and code to help troubleshoot the problem.
Yellow= 5mm Red=7mm margin Troubleshooting your system 8 What it means There is a problem with the meter. What to do Do Not use the meter. Contact Customer Service. Contact OneTouch® Customer Care on 0800 121 200 (UK) or 1800 535 676 (Ireland). What it means Error message could be caused either by a used test strip or a problem with the meter or test strip. What to do Repeat the test with a new test strip; see page 48 or page 63. If this message continues to appear, contact Customer Service.
Yellow= 5mm Red=7mm margin 8 Troubleshooting your system What it means The sample was applied before the meter was ready. What to do Repeat the test with a new test strip. Apply a blood or control solution sample only after the flashing symbol appears on the display. If this message continues to appear, contact Customer Service. Contact OneTouch® Customer Care on 0800 121 200 (UK) or 1800 535 676 (Ireland). 05983201A_OT_UPF_OB_mmolL_Inside_R2.
Yellow= 5mm Red=7mm margin Troubleshooting your system 8 What it means One of the following may apply: • Not enough blood or control solution was applied or more was added after the meter began to count down. • The test strip may have been damaged or moved during testing. • The sample was improperly applied. • There may be a problem with the meter. What to do Repeat the test with a new test strip; see page 48 or page 63. If the error message appears again, contact Customer Service.
Yellow= 5mm Red=7mm margin 8 Troubleshooting your system What it means The meter has detected a problem with the test strip. Possible cause is test strip damage. What to do Repeat the test with a new test strip; see page 48 or page 63. If the error message appears again, contact Customer Service. Contact OneTouch® Customer Care on 0800 121 200 (UK) or 1800 535 676 (Ireland). What it means Meter battery power is low but there is still enough battery power to perform a test.
Yellow= 5mm Red=7mm margin Troubleshooting your system 8 What it means There is not enough battery power to perform a test. What to do Replace the battery immediately. What it means No result in memory, such as the first time use or when you have chosen to delete all the results after downloading to a computer. What to do Contact Customer Service to report this occurrence, unless this is your first use of the meter.
Yellow= 5mm Red=7mm margin 9 Detailed information about your system Comparing meter results to laboratory results Results obtained from the OneTouch Ultra Plus Flex™ Meter and laboratory tests are reported in plasmaequivalent units. However, your meter result may differ from your lab result due to normal variation.
Yellow= 5mm Red=7mm margin Detailed information about your system 9 Meter results can be affected by factors that do not affect lab results in the same way. Specific factors that may cause your meter result to vary from your lab result may include: • You have eaten recently. This can cause a result from fingertip testing to be up to 3.9 mmol/L higher than a lab test using blood drawn from a vein.1 • Your hematocrit is above 60% or below 20%. • You are severely dehydrated.
Yellow= 5mm Red=7mm margin 9 Detailed information about your system Guidelines for obtaining an accurate meter to lab comparison Before going to the lab: • Perform a control solution test to make sure your meter is working properly. • Do Not eat for at least 8 hours before you test your blood. • Take your meter and testing supplies with you to the lab. 05983201A_OT_UPF_OB_mmolL_Inside_R2.
Yellow= 5mm Red=7mm margin Detailed information about your system 9 Testing with your OneTouch Ultra Plus Flex™ Meter at the lab: • Test within 15 minutes of the lab test. • Use only a fresh, capillary blood sample from your fingertip. • Follow all instructions in this Owner's Booklet for performing a blood glucose test.
Yellow= 5mm Red=7mm margin 9 Detailed information about your system Technical specifications Assay method FAD-GDH (flavin adenine dinucleotide dependent glucose dehydrogenase) Automatic shutoff Two minutes after last action Battery ratings 3.0 V d.c. (CR2032 lithium coin cell battery), Battery type One replaceable 3.0 Volt CR2032 lithium coin cell battery (or equivalent) BLUETOOTH® SMART Technology Frequency range: 2.4-2.
Yellow= 5mm Red=7mm margin Detailed information about your system 9 Memory 500 test results Operating ranges Temperature: 6-44°C Relative humidity: non-condensing 10-90% Altitude: up to 3048 meters Hematocrit: 20-60% Reported result range 1.1-33.3 mmol/L Sample Fresh capillary whole blood Sample volume 0.4 µL Size 52.0(W) x 86.0(L) x 16.0(T) mm Test time 5 seconds Unit of measure mmol/L Weight Approximately 50 grams 05983201A_OT_UPF_OB_mmolL_Inside_R2.
Yellow= 5mm Red=7mm margin 9 Detailed information about your system System Accuracy Diabetes experts have suggested that glucose meters should agree within TBD mmol/L of a laboratory method when the glucose concentration is lower than TBD mmol/L, and within TBD of a laboratory method when the glucose concentration is TBD mmol/L or higher. Samples from TBD patients were tested using both the OneTouch Ultra Plus Flex™ System and the YSI 2300 Glucose Analyzer laboratory instrument.
Yellow= 5mm Red=7mm margin Detailed information about your system 9 System Accuracy Results for Glucose Concentrations ≥TBD mmol/L Percent (and number) of meter results that match the laboratory test Within ±TBD Within ±TBD Within ±TBD TBD TBD TBD System Accuracy Results for Glucose Concentrations between TBD mmol/L and TBD mmol/L Within ±TBD mmol/L or ±TBD TBD NOTE: Where TBD mmol/L represents the lowest glucose reference value and TBD mmol/L represents the highest glucose reference value (YSI val
Yellow= 5mm Red=7mm margin 9 Detailed information about your system Regression Statistics Samples were tested in duplicate on each of three test strip lots. Results indicate that the OneTouch Ultra Plus Flex™ System compares well with a laboratory method. # of Subjects # of Tests Slope Intercept (mmol/L) TBD TBD TBD TBD 95% CI Slope 95% CI Intercept (mmol/L) Std. Error (Sy.
Yellow= 5mm Red=7mm margin Detailed information about your system 9 Precision Within Run Precision (TBD Venous Blood Samples Tested per Glucose Level) Data generated using the OneTouch Ultra Plus Flex™ Meter.
Yellow= 5mm Red=7mm margin 9 Detailed information about your system Total Precision (TBD Control Solution Tests per Glucose Level) Data generated using the OneTouch Ultra Plus Flex™ Meter. Glucose Level Ranges (mmol/L) Mean Glucose (mmol/L) Standard Deviation (mmol/L) Coefficient of Variation (%) Low (TBD) TBD TBD TBD Mid (TBD) TBD TBD TBD High (TBD) TBD TBD TBD 05983201A_OT_UPF_OB_mmolL_Inside_R2.
Yellow= 5mm Red=7mm margin Detailed information about your system 9 Guarantee LifeScan guarantees that the OneTouch Ultra Plus Flex™ Meter will be free of defects in material and workmanship for three years, valid from the date of purchase. The guarantee extends only to the original purchaser and is not transferable. Electrical and safety standards This meter complies with CISPR 11:Class B (Radiated Only).
Yellow= 5mm Red=7mm margin 10 Index AST 39 Batteries 82 Batteries, replacing 82 Battery empty icon 3 Battery empty warning 95 Battery low icon 3, 94 BLUETOOTH SMART feature 9, 12, 27, 69 BLUETOOTH® SMART feature 7 Buttons on meter 18 Cleaning your meter, lancing device and cap 78 Comparing meter results to laboratory results 96 Compatible wireless devices 3 Control solution 60 Control solution, discard and expiration dates 58, 59 Control solution, testing 58, 64, 66 Data port 18 Date setting 24, 26 Dehyd
Yellow= 5mm Red=7mm margin Index 10 Edit time Error messages EXTREME HIGH GLUCOSE message EXTREME LOW GLUCOSE message Fingertip testing procedure First time setup Guarantee Hyperglycemia Hypoglycemia Icons Infection, reduce the chance Intended use Kit components Lancing device mmol/L Pairing PIN Plasma calibration Range Indicator feature Range limits setting Results, reviewing past Serial number Settings Software, diabetes management 05983201A_OT_UPF_OB_mmolL_Inside_R2.
Yellow= 5mm Red=7mm margin 10 Index Start-up screen 22 Storing your system 78 Symbols 3, 4 Syncing 32, 69 Technical specifications 102 Temperature 34, 66, 78, 88, 89, 101 Test principle 7 Test strip 19, 45, 60 Test strip, applying drop of blood 48 Test strip, discard and expiration dates 35 Test strip, silver prongs 19 Time setting 23, 26 Turning meter off 33 Turning meter on 22, 45, 60 Unexpected results 52, 54 Unit of measure 18, 101 Unusual red blood cell count (hematocrit) 97, 101 05983201A_OT_UPF_OB
Yellow= 5mm Red=7mm margin Notes 05983201A_OT_UPF_OB_mmolL_Inside_R2.
Yellow= 5mm Red=7mm margin Notes 05983201A_OT_UPF_OB_mmolL_Inside_R2.
05983201A_OT_UPF_OB_mmolL_Inside_R2.indd 1 1/27/15 4:35 PM Template, OneTouch Ultra Plus Flex, Owner's Booklet, 1L GC 059832 N/A CMYK ADS Reference Number: ADS-03 / LCB-01a GC Document Number: GC AW Template Number: AW 05983201A GC AW Template Title: 01/27/15 28696 Number of Colors: 4 Job No: LifeScan Contact: R. Muolic Rev.