Operation Manual

3
GETTING TO KNOW YOUR SYSTEM
Turning your meter on
To perform a test, insert a test strip as far as it will go. The display will turn on and
the meter will briey perform system checks. Or, to change the time and date,
start with the meter o, then press and hold for ve seconds until the start-up
test screen appears. After the start-up test screen, the pre-set time and date will
appear on the display. Or, if you want to turn the meter on to review past results,
start with the meter o, then press and release .
Every time you turn your meter on, a start-up test screen will appear for two seconds. All segments of the display should appear
briey on the start-up test screen to tell you that the meter is working properly. To check that all display segments are working, as
soon as the start-up test screen appears, press and hold to keep the start-up test screen display on. Release to proceed to the
next step. If the meter does not power on, try changing the meter battery. See Replacing the battery in Section 6.
Turning your meter o
There are several ways to turn your meter o:
• Press and hold for two seconds, when reviewing past results.
• Your meter will turn o by itself if left alone for two minutes.
• Before or after completing a test, remove the test strip.
CAUTION: If any information is missing from the
start-up test screen, there may be a problem with the
meter. Call OneTouch® Customer Care 0800 121200 (UK)
or 1800 535676 (Ireland).
Start-up Test Screen
Downloaded from www.Manualslib.com manuals search engine