Operation Manual

19
CONTROL SOLUTION TESTING
Press so that the control solution test symbol CtL appears in the upper right corner of the
display. You must mark the test before you apply control solution. Once you have completed
the test, you cannot change the marking. The meter is now ready to perform a control solution
test. If you decide not to do a control solution test, press again to remove CtL from
the display.
Prepare and apply control solution
Shake the control solution vial before each test. Remove the cap and squeeze the vial to
discard the rst drop. Then wipe the tip with a clean tissue or cloth. Hold the vial upside
down and gently squeeze a hanging drop. Touch and hold the hanging drop of control
solution to the narrow channel in the top edge of the test strip. Make sure the conrmation
window lls completely. Control solution should not be applied to the at face of the
test strip.
Read your result
When the conrmation window is full, the meter will count down from 5 to 1.
Your result will then appear on the display, along with CtL and the unit of measurement.
Check if the result is in range
Compare the result displayed on the meter to the control solution range printed on the test strip vial. Each vial of
test strips may have a dierent control solution range. If the results you get are not within this range, the meter and
test strips may not be working properly. Repeat the control solution test.
Out-of-range results may be due to:
• not following the instructions detailed in steps 1–7,
• expired or contaminated control solution,
• expired or damaged test strip,
• use of a test strip or control solution past its discard date, or
• a problem with the meter.
CAUTION: The control solution range printed on the test strip vial is for OneTouch® Ultra® Control Solution only. It is not a
recommended range for your blood glucose level.
CAUTION: If you continue to get control solution results that fall outside the range printed on the test strip vial, Do Not use the meter,
the test strips, or the control solution. Call OneTouch® Customer Care 0800 121200 (UK) or 1800 535676 (Ireland).
Conrmation
Window
5.5–7.4 mmol/L
(Example)
Top
Edge
Narrow Channel
Apply drop here