Operation Manual

Detailed information
about your system
9
135
Total Precision (600 Control Solution Tests per Glucose
Level)
Data generated using the OneTouch Select® Plus Meter.
Glucose
Level
Ranges
(mmol/L)
Mean
Glucose
(mmol/L)
Standard
Deviation
(mmol/L)
Coecient
of Variation
(%)
Low
(1.67-2.78)
2.61 0.06 2.49
Mid
(5.33-7.99)
6.15 0.11 1.83
High
(15.54-23.31)
20.40 0.31 1.52