Precision
Within Run Precision (300 Venous Blood Samples Tested
per Glucose Level)
Data generated using the OneTouch Select® Plus Meter.
Target
Glucose
(mmol/L)
Mean
Glucose
(mmol/L)
Standard
Deviation
(mmol/L)
Coecient
of Variation
(%)
2.2 2.50 0.09 3.59
3.6 3.90 0.10 2.56
6.7 7.06 0.16 2.25
11.1 11.57 0.29 2.52
19.4 19.55 0.40 2.07
Results show that the greatest variability observed
between test strips when tested with blood is 0.10 mmol/L
SD or less at glucose levels less than 5.55 mmol/L, or
2.52% CV or less at glucose levels at 5.55 mmol/L or