System Accuracy Results for Glucose
Concentrations ≥5.55 mmol/L
Percent (and number) of meter results that match the
laboratory test
Within ±5% Within ±10% Within ±15%
72.7%
(314/432)
98.6%
(426/432)
100.0%
(432/432)
System Accuracy Results for Glucose
Concentrations between 1.2 mmol/L
and 25.9 mmol/L
Within ±0.83 mmol/L or ±15%
99.5% (597/600)
NOTE: Where 1.2mmol/L represents the lowest glucose
reference value and 25.9mmol/L represents the highest
glucose reference value (YSI value).