Operation Manual

8
Troubleshooting
120
There are two Error 6 screens:
What it means
The meter has detected a
problem with the test strip.
Possible cause is that you did
not apply a blood sample to the
test strip when prompted by the
meter.
What to do
Repeat the test with a new
test strip; see page32. If
the error message appears
again, contact Customer Care.
Contact OneTouch® Customer
Care on 0800121200 (UK)
or 1800535 676 (Ireland)
80062253 (Malta) or 800 9074
(Iceland).
Error 6
Perform blood
test again with
a new strip.