Operation Manual
112
Detailed information about your system
9
Total Precision (600 Control Solution Tests
per Glucose Level)
Data generated using the OneTouchSelect Plus Flex™
Meter.
Glucose Level
Ranges
(mg/dL)
Mean
Glucose
(mg/dL)
Standard
Deviation
(mg/dL)
Coecient
of Variation
(%)
Low
(30-50)
45.45 1.23 2.71
Mid
(96-144)
110.44 2.01 1.82
High
(280-420)
365.84 6.04 1.65