Operation Manual

107
Detailed information about your system
9
System Accuracy
Diabetes experts have suggested that glucose meters
should agree within 15mg/dL of a laboratory method
when the glucose concentration is lower than 100mg/dL,
and within 15% of a laboratory method when the glucose
concentration is 100mg/dL or higher. Samples from 100
patients were tested using both the OneTouchSelect
Plus Flex™ System and the YSI 2300 Glucose Analyzer
laboratory instrument.
System Accuracy Results for Glucose
Concentrations <100mg/dL
Percent (and number) of meter results that match the
laboratory test
Within
±5mg/dL
Within
±10mg/dL
Within
±15mg/dL
67.2%
(121/180)
93.3%
(168/180)
96.7%
(174/180)