(1) SET Reader function (SETR)
This command writes the functional conditions of the command in process into the nonvolatile memory in
the Reader/ Writer. The setting shall be enabled at the upcoming startup.
Command SETR (S) param1=value1 (S)...param n=value n<LF>
It specifies a parameter connecting a setting item to a setting value with equal.
(The setting item does not have to be prepended with ‘ -‘.)
No space shall be put in from the beginning of the setting item to the end of the
setting value. If the setting value includes any space, the space shall be enclosed with
double quotation.
In a single command, multiple setting items can be specified by separating with
space. A 2 or more character space is acceptable for indicating separation.
Response SETR0000<LF>
*(S) Space Character
Example 1) When DHCP Enabled is set “ON”:
[Tx] SETR dhe=1<LF>
[Rx] SETR0000<LF>
Example 2) When [Host name; V750-BA50C04, IP; 192.168.1.1, and Net Mask; 255.255.255.0] is set:
[Tx] SETR lhn=“V750 BA50C04” lip=192.168.1.1 lnm=255.255.255.0<LF>
[Rx] SETR0000<LF>
Example 3) When [RS-232C Baud rate; 19200bps, Data length; 7bit, Parity; Even, and Stop; 1bit] is
set:
[Tx] SETR rbr=19200 rdt=7 rpb=E rsb=1<LF>
[Rx] SETR0000<LF>
Example 4) When the setting item is incorrect:
[Tx] SETR xxx=111<LF>
[Rx] SETR1407<LF>