Operation Manual

24
The particle size is measured according to the provisions of the EN13544-1:2007 “Respiratory therapy
equipment - Part 1: Nebulising systems and their components”, Annex CC.3 using the multistage
cascade impactor method.
Particle Size:
*MMAD 4.2 µm
MMAD = Mass Median Aerodynamic Diameter
Medication Container Capacity:
7 ml maximum
Sound:
Less than 20 dB
Nebulisation rate:
**Greater than 0.25 ml/min (by weight loss)
Aerosol Output:
*0.69 ml (2ml, 1%NaF)
Aerosol Output Rate:
*0.087 ml/min (2ml, 1%NaF)
Notes:
The distribution of particle size and aerosol output are measured with 2.5% NaF solution. The particle size
distribution and aerosol output may vary by combination of product, medication and ambient conditions such
as temperature, humidity and atmospheric pressure.
Performance may vary with drugs such as suspensions or high viscosity. See drug supplier’s data sheet for
further details.
See web site of OMRON HEALTHCARE EUROPE to update technical information.
URL: www.omron-healthcare.com
* Independently measured at SolAero Ltd., Canada, Dr. John Dennis, according to EN13544-1:2007.
** The Nebulisation rate is measured with saline 0.9% solution at 23°C and can vary with medication and
ambient conditions.
Cumulative % particle mass of sodium fluoride under size
0
10
20
30
40
50
60
70
80
90
100
0010111.0
Particle Size Dp (μm)
Cumulative Undersize %
Individual
Mean
MMAD 4.2 μm
Technical data