HEM-7119-E(V)_A_M04_110812.pdf HEM-7119-E(V)_A_M.
HEM-7119-E(V)_main.book Page 2 Thursday, September 29, 2011 11:48 AM Contents Before using the unit Introduction .................................................................................. 3 Important Safety Information...................................................... 4 1. Overview .................................................................................. 7 2. Preparation .............................................................................. 9 2.
HEM-7119-E(V)_main.book Page 3 Thursday, September 29, 2011 11:48 AM Before using the unit Introduction Thank you for purchasing the OMRON M2 Upper Arm Blood Pressure Monitor. The OMRON M2 is a fully automatic blood pressure monitor, operating on the oscillometric principle. It measures your blood pressure and pulse rate simply and quickly. For comfortable controlled inflation without the need of pressure pre-setting or re-inflation the device uses its advanced “IntelliSense” technology.
HEM-7119-E(V)_main.book Page 4 Thursday, September 29, 2011 11:48 AM Important Safety Information Important Safety Information Consult your doctor prior to using in pregnancy or if diagnosed with arrhythmia or arteriosclerosis. Please read this section carefully before using the unit. Warning: • Indicates a potentially hazardous situation which, if not avoided, could result in death or serious injury. (General Usage) • Always consult your doctor.
HEM-7119-E(V)_main.book Page 5 Thursday, September 29, 2011 11:48 AM Important Safety Information • Do not leave the cuff wrapped on the arm if taking measurements during the night. This could result in injury. • Do not inflate the arm cuff over 299 mmHg. • Do not use a mobile phone or other devices that emit electromagnetic fields, near the unit. This may result in incorrect operation of the unit. • Do not operate the unit in a moving vehicle (car, airplane).
HEM-7119-E(V)_main.book Page 6 Thursday, September 29, 2011 11:48 AM Important Safety Information • Do not wash the arm cuff or immerse it in water. • Read and follow the “Important information regarding Electro Magnetic Compatibility (EMC)” in the Technical Data Section. • Read and follow the “Correct Disposal of This Product” in the Technical Data Section when disposing of the device and any used accessories or optional parts.
HEM-7119-E(V)_main.book Page 7 Thursday, September 29, 2011 11:48 AM 1.Overview 1. Overview Main Unit A E D F B A. B. C. D. C Display O/I START button M (Memory) button AC adapter jack (for optional AC adapter) E. Battery compartment F. Air jack EN Arm Cuff G H I G. Arm cuff (Medium cuff: arm circumference 22-32 cm) H. Air tube I.
HEM-7119-E(V)_main.book Page 8 Thursday, September 29, 2011 11:48 AM 1.Overview Display J O K L P M Q R N S J. Systolic blood pressure K. Diastolic blood pressure L. Heartbeat symbol (Flashes during measurement) M. Deflation symbol N. Battery low symbol O. P. Q. R. S. Blood pressure level indicator Pulse display Cuff wrapping guide Memory symbol Irregular heartbeat symbol Package Contents (others) T T. Storage case U.
HEM-7119-E(V)_main.book Page 9 Thursday, September 29, 2011 11:48 AM 2.Preparation 2. Preparation Installing/Replacing the Batteries 2.1 1. Turn the main unit upside down. 2. Slide the battery cover in the direction of the arrow while pressing the ribbed part of the cover. 3. Install or replace four “AAA” size batteries so that the + (positive) and - (negative) polarities match the polarities indicated on the battery compartment. 4. Put the battery cover back in place.
HEM-7119-E(V)_main.book Page 10 Thursday, September 29, 2011 11:48 AM 2.Preparation Battery Life & Replacement If the battery low symbol ( ) appears on the display, replace all four batteries at the same time. - When the battery low symbol ( ) starts to blink, you will still be able to use the unit for a short while. You should replace the batteries with new ones as soon as possible. - When the symbol ( ) remains lit, the batteries are exhausted. You should replace the batteries with new ones at once.
HEM-7119-E(V)_main.book Page 11 Thursday, September 29, 2011 11:48 AM Operating instructions 3.Using the Unit 3. Using the Unit 3.1 How to Sit Correctly When Taking a Measurement Correct posture during measurement is necessary to get accurate results. Notes: • Measurements should be taken in a quiet place and you should be in a relaxed, seated position. Make sure that the room is not too hot or too cold.
HEM-7119-E(V)_main.book Page 12 Thursday, September 29, 2011 11:48 AM 3.Using the Unit Note: You can take a measurement on either your left or right arm. The blood pressure can differ between the right arm and the left arm and therefore also the measured blood pressure values can be different. OMRON recommends to always use the same arm for measurement. If the values between the two arms differ substantially, please check with your physician which arm to use for your measurement.
HEM-7119-E(V)_main.book Page 13 Thursday, September 29, 2011 11:48 AM 3.Using the Unit 3.2 Applying the Arm Cuff Be sure to wrap the arm cuff correctly so that you get accurate results. Measurements can be taken in light clothing. However, please remove thick clothes, such as sweaters, before taking a reading. 1. Insert the air plug into the air jack on the left side of the main unit. 2. Put your left arm through the cuff loop.
HEM-7119-E(V)_main.book Page 14 Thursday, September 29, 2011 11:48 AM 3.Using the Unit 3. Position the arm correctly. 1) The air tube should run down the inside of your forearm and be in line with your middle finger. 2) The bottom of the cuff should be approximately 1 to 2 cm above your elbow. 3) Apply the cuff to your upper arm so that the marker (arrow under tube) is centered on the middle of your inner arm and points down the inside of the arm.
HEM-7119-E(V)_main.book Page 15 Thursday, September 29, 2011 11:48 AM 3.Using the Unit 3.3 Taking a Reading 1. Press the O/I START button to turn the unit on. The cuff starts to inflate automatically. Sit in a chair with your feet flat on the floor and place your arm on a table so that the cuff will be at the same level as your heart. Keep still and do not talk during measurement. Note: To cancel a measurement, press the O/I START button to turn the unit off and to release the air in the arm cuff.
HEM-7119-E(V)_main.book Page 16 Thursday, September 29, 2011 11:48 AM 3.Using the Unit 4. Check the measurement results. The unit automatically stores blood pressure and pulse rate into its memory. Refer to “3.5 Using the Memory Function”. Systolic blood pressure Diastolic blood pressure Pulse display Note: Wait 2-3 minutes before taking another blood pressure measurement. Waiting between readings allows the arteries to return to the condition prior to taking the blood pressure measurement.
HEM-7119-E(V)_main.book Page 17 Thursday, September 29, 2011 11:48 AM 3.Using the Unit 5. Remove the arm cuff. 6. Press the O/I START button to turn the unit off. Note: If you forget to turn the unit off, it will automatically shut itself off after five minutes.
HEM-7119-E(V)_main.book Page 18 Thursday, September 29, 2011 11:48 AM 3.Using the Unit Important: Recent research suggests that the following values can be used as a guide to high blood pressure for measurements taken at home. Systolic Blood Pressure Above 135 mmHg Diastolic Blood Pressure Above 85 mmHg This criteria is for home blood pressure measurement. For professional office blood pressure measurement criteria, please refer to Chapter 9 “Some Useful Information about Blood Pressure”.
HEM-7119-E(V)_main.book Page 19 Thursday, September 29, 2011 11:48 AM 3.Using the Unit What is Irregular Heartbeat? An irregular Normal Heartbeat heartbeat is a Pulse heartbeat rhythm that varies by more Blood pressure than 25% from the Irregular Heartbeat average heartbeat Short Long rhythm detected Pulse while the unit is measuring the Blood pressure systolic and diastolic blood pressure.
HEM-7119-E(V)_main.book Page 20 Thursday, September 29, 2011 11:48 AM 3.Using the Unit 3.4 Instructions for Special Conditions If your systolic pressure is known to be more than 220 mmHg, press and hold the O/I START button until the arm cuff inflates 30 to 40 mmHg higher than your suspected systolic pressure. 1. Press the O/I START button to turn the unit on. Measurement starts. 2.
HEM-7119-E(V)_main.book Page 21 Thursday, September 29, 2011 11:48 AM 3.Using the Unit 3.5 Using the Memory Function The unit automatically stores up to 21 sets of measurement values (blood pressure and pulse rate). When 21 sets of measurement values are stored, the oldest record is deleted to save the most recent values. 1. Press the M button. Note: If there are no measurements results stored in memory, the screen to the right is displayed. 2.
HEM-7119-E(V)_main.book Page 22 Thursday, September 29, 2011 11:48 AM 3.Using the Unit 3. Press the O/I START button to turn the unit off. Note: If you forget to turn the unit off, it will automatically shut itself off after five minutes. To Delete All the Values Stored in Memory You cannot delete individual stored readings, all the readings in the unit will be deleted. 1. To delete stored readings, first press the M button.
HEM-7119-E(V)_main.book Page 23 Thursday, September 29, 2011 11:48 AM Care and maintenance 4.Quick Reference Guide 4. Quick Reference Guide If you are using this device for the first time, please read carefully chapter 3 of this instruction manual. To help ensure a reliable reading, avoid eating, drinking alcohol, smoking, or exercising for at least 30 minutes before taking a measurement. Note: Remove any tight-fitting clothing from your upper arm. 1.
HEM-7119-E(V)_main.book Page 24 Thursday, September 29, 2011 11:48 AM 5.Handling Errors and Problems 5. Handling Errors and Problems 5.1 Error Messages Error Display 24 Cause Remedy Cuff is under inflated. Carefully read and repeat the steps listed under section 3.4. Movement during measurement. Repeat measurement. Remain still and do not talk during measurement. Refer to section 3.3. Air plug disconnected. Insert the air plug securely. Refer to section 3.2. Arm cuff not applied correctly.
HEM-7119-E(V)_A_M.fm Page 25 Monday, October 10, 2011 8:48 AM 5.Handling Errors and Problems Error Display Cause Remedy Battery power is low. Replace all four “AAA” batteries with new ones. Refer to section 2.1. Device error. Contact your OMRON retail outlet or distributor. Blinks or appears continuously Note: The irregular heartbeat symbol ( with error messages.
HEM-7119-E(V)_main.book Page 26 Thursday, September 29, 2011 11:48 AM 5.Handling Errors and Problems 5.2 Troubleshooting Problem Remedy Arm cuff not applied correctly. Apply the arm cuff correctly. Refer to section 3.2. Movement or talking during measurement. Remain still and do not talk during measurement. Refer to section 3.3. Clothing is interfering with the arm cuff. Remove any clothing interfering with the arm cuff. Refer to section 3.1.
HEM-7119-E(V)_main.book Page 27 Thursday, September 29, 2011 11:48 AM 5.Handling Errors and Problems Problem Nothing happens when you press the buttons. Other problems. Cause Remedy The batteries are empty. Replace the batteries with new ones. The batteries have been inserted incorrectly. Insert the batteries with the correct (+/-) polarity. Press the O/I START button and repeat measurement. If the problem continues, try replacing the batteries with new ones.
HEM-7119-E(V)_A_M.fm Page 28 Monday, October 10, 2011 8:49 AM 6.Maintenance and Storage 6. Maintenance and Storage Maintenance To protect your unit from damage, please observe the following: • Do not subject the main unit and cuff to extreme temperatures, humidity, moisture or direct sunlight. • Do not fold the cuff or tubing tightly. • Do not disassemble the unit. • Do not subject the unit to strong shocks or vibrations (for example, dropping the unit on the floor).
HEM-7119-E(V)_main.book Page 29 Thursday, September 29, 2011 11:48 AM 6.Maintenance and Storage Storage Keep the unit in its storage case when not in use. 1. Unplug the air tube from the air jack. Gently fold the air tube into the arm cuff. Note: Do not bend the air tube excessively. 2. Place the arm cuff and main unit in the storage case. Do not store the unit in the following situations: • If the unit is wet.
HEM-7119-E(V)_main.book Page 30 Thursday, September 29, 2011 11:48 AM 7.Optional Parts 7. Optional Parts Wide Range Cuff Arm circumference 22 - 42 cm CW-9520534-2 (Model: HEM-RML30) Large Arm Cuff Arm circumference 32 - 42 cm Medium Arm Cuff Arm circumference 22 - 32 cm CM2-9513256-6 AC Adapter S CL2-9513255-8 Adapter S-9515336-9 Using the Optional AC Adapter Insert batteries into the battery compartment even when using the AC adapter. Notes: • Never plug or unplug the power cord with wet hands.
HEM-7119-E(V)_main.book Page 31 Thursday, September 29, 2011 11:48 AM 8.Technical Data 8.
HEM-7119-E(V)_main.book Page 32 Thursday, September 29, 2011 11:48 AM 8.Technical Data Important information regarding Electro Magnetic Compatibility (EMC) With the increased number of electronic devices such as PC’s and mobile (cellular) telephones, medical devices in use may be susceptible to electromagnetic interference from other devices. Electromagnetic interference may result in incorrect operation of the medical device and create a potentially unsafe situation.
HEM-7119-E(V)_main.book Page 33 Thursday, September 29, 2011 11:48 AM 9.Some Useful Information about Blood Pressure 9. Some Useful Information about Blood Pressure What is Blood Pressure? Blood pressure is a measure of the force of blood flowing against the walls of the arteries. Arterial blood pressure is constantly changing during the course of the heart’s cycle. The highest pressure in the cycle is called the Systolic Blood Pressure; the lowest is the Diastolic Blood Pressure.
HEM-7119-E(V)_main.book Page 34 Thursday, September 29, 2011 11:48 AM 9.Some Useful Information about Blood Pressure Classification of Blood Pressure by the World Health Organization The World Health Organization (WHO) and the International Society of Hypertension (ISH) developed the Blood Pressure Classification shown in this figure.
HEM-7119-E(V)_main.book Page 35 Thursday, September 29, 2011 11:48 AM Manufacturer OMRON HEALTHCARE Co., Ltd. 53, Kunotsubo, Terado-cho, Muko, Kyoto, 617-0002 JAPAN EU-representative OMRON HEALTHCARE EUROPE B.V. Scorpius 33, 2132 LR Hoofddorp THE NETHERLANDS www.omron-healthcare.com Production facility OMRON HEALTHCARE MANUFACTURING VIETNAM CO., LTD. No.