NE-C801S-KDE_A_M_main.book Page 2 Friday, January 13, 2012 4:43 PM QUICK MANUAL Please read entire manual before using the device. Warning: Clean and disinfect all the parts before using them. page 16-18 1. Adding medication Rotate the inhalation top anti-clockwise to remove. Add the appropriate amount of prescribed medication. Make sure the vaporiser head is securely attached to the inhalation top. Clicks Put the inhalation top back and rotate it clockwise until they click in place.
NE-C801S-KDE_A_M_main.book Page 3 Friday, January 13, 2012 4:43 PM 3. Connecting AC adapter Power Outlet Power Plug Power Connector Insert the power plug of AC adapter into the power connector of the main unit. Plug the AC adapter into the power outlet. page 11 4. Attaching air tube Attach the air tube to the main unit and to the nebuliser kit. When do so, twist and push it firmly. page 13 5. Applying mask and starting inhalation Apply the mask securely. Turn on the power to the main unit.
NE-C801S-KDE_A_M_main.book Page 4 Friday, January 13, 2012 4:43 PM Before using the device Contents Before using the device Introduction ................................................................................................................. 4 Intended use ................................................................................................................ 5 Important safety instructions..................................................................................... 6 1.
NE-C801S-KDE_A_M_main.book Page 5 Friday, January 13, 2012 4:43 PM Intended use Medical Purpose Intended User Intended Patients Environment Durable period Precautions for use This product is intended to be used for inhaling medication for respiratory disorders. • Legally certified medical experts, such as doctor, nurse and therapist. • Caregiver or patient under the guidance of qualified medical experts for home treatment.
NE-C801S-KDE_A_M_main.book Page 6 Friday, January 13, 2012 4:43 PM Important safety instructions Read all the information in the instruction manual and any other literature included in the box before using the device. When using an electrical product, especially when children are present, basic safety precautions should always be followed including the following: Warning: Indicates a potentially hazardous situation which, if not avoided, could result in serious injury.
NE-C801S-KDE_A_M_main.book Page 7 Friday, January 13, 2012 4:43 PM Caution: Indicates a potentially hazardous situation which if not avoided, may result in minor or moderate injury, or physical damage. (Usage) • Provide close supervision when this device is used by, on, or near children or invalids. • Make sure that the parts are attached correctly. • Make sure that the vaporiser head is correctly installed before use. • Make sure that the air filter is correctly attached.
NE-C801S-KDE_A_M_main.book Page 8 Friday, January 13, 2012 4:43 PM (Risk of electrical shock) • Always unplug the AC adapter from the power outlet after use and before cleaning. • Plug the device into the appropriate voltage outlet. Do not overload power outlets or use extension cords. • Always remove the AC adapter from the device after use and before cleaning. • Do not misuse the cord of the AC adapter. • Do not wind the cord of the AC adapter around the compressor or AC adapter.
NE-C801S-KDE_A_M_main.book Page 9 Friday, January 13, 2012 4:43 PM 1.
NE-C801S-KDE_A_M_main.book Page 10 Friday, January 13, 2012 4:43 PM 1.
NE-C801S-KDE_A_M_main.book Page 11 Friday, January 13, 2012 4:43 PM Operating instructions 2. Preparing the nebuliser for use Warning: • Clean and disinfect the nebuliser kit, kids accessory, mask or mouthpiece before using them for the first time after purchase. • If the device has not been used for a long period of time, or if more than one person uses the same device, clean and disinfect the nebuliser kit, kids accessory, mask or mouthpiece before using them.
NE-C801S-KDE_A_M_main.book Page 12 Friday, January 13, 2012 4:43 PM 2.Preparing the nebuliser for use 2.2 Adding medication and assembling the parts 1. Remove the mask or mouthpiece, kids accessory and cap from the nebuliser kit. 2. Remove the inhalation top from the medication tank. 1) Rotate the inhalation top anti-clockwise. 2) Lift the inhalation top out of the medication tank. 3. Add the correct amount of prescribed medication to the medication tank.
NE-C801S-KDE_A_M_main.book Page 13 Friday, January 13, 2012 4:43 PM 2.Preparing the nebuliser for use 6. Attach the cap, kids accessory, mask or mouthpiece to the nebuliser kit. 1) Attach the cap to the inhalation air inlet. 2) Attach the kids accessory to the inhalation top. 3) Attach the mask or mouthpiece to the inhalation top. Note: If the amount of aerosol is too small, it is not necessary to place the cap. 7. Attach the air tube.
NE-C801S-KDE_A_M_main.book Page 14 Friday, January 13, 2012 4:43 PM 3. Taking medication 1. Hold the nebuliser kit as indicated on the right. Caution: Do not tilt the nebuliser kit so that it is at an angle of greater than 45° in all directions. Medication may flow into the mouth or it may result in not nebulising effectively. Right angle 45q 45q 2. Press the power switch to the on ( ) position. The compressor starts and nebulisation starts. Note: Observe whether aerosol is generated.
NE-C801S-KDE_A_M_main.book Page 15 Friday, January 13, 2012 4:43 PM 3.Taking medication 4. When treatment is complete, turn the power off. Check that there is no condensation or moisture in the air tube and unplug the device from the power outlet. Warning: Condensation may occur in the air tube. Do not store the air tube while it contains condensation or moisture. This could result in infection via bacteria.
NE-C801S-KDE_A_M_main.book Page 16 Friday, January 13, 2012 4:43 PM Care and maintenance 4. Cleaning and daily disinfecting 4.1 Cleaning Following the cleaning instructions after each use will prevent any remaining medication in the tank from drying, resulting in the device not nebulising effectively and also for preventing infections.
NE-C801S-KDE_A_M_main.book Page 17 Friday, January 13, 2012 4:43 PM 4.Cleaning and daily disinfecting 4.2 Disinfecting the nebulising parts Always clean before disinfection. (See Section 4.1 “Cleaning”) Always disinfect the medication tank, mouthpiece and other nebulising parts after the last treatment of the day. For appropriate method of disinfecting each part, refer to page 18. To disinfect the parts, use one of the methods described below: A. Use a commercially available disinfectant.
NE-C801S-KDE_A_M_main.book Page 18 Friday, January 13, 2012 4:43 PM 4.Cleaning and daily disinfecting Disinfecting ethanol Alcohol O Milton* (0.1%) Sodium hypochlorite O O Osvan* (0.1%) Quaternary ammonium O O O Hibitane* (0.5%) Chlorhexidine × O O O Tego* (0.
NE-C801S-KDE_A_M_main.book Page 19 Friday, January 13, 2012 4:43 PM 5. Maintenance and storage To keep your device in the best condition and protect the device from damage follow these directions: Caution: • Do not leave the device unattended with infants or persons who cannot express their consent. • Do not subject the device or the parts to any strong shocks such as dropping the device on the floor. • Do not carry or leave the nebuliser kit while the medication tank contains medication.
NE-C801S-KDE_A_M_main.book Page 20 Friday, January 13, 2012 4:43 PM 6. Changing the air filter If the air filter has changed colour, or has been used on average for more than 60 days, replace it with a new one. 1. Pull the air filter cover off the compressor. 2. Change the air filter. 3. Put the air filter cover back in place. Notes: • Use only OMRON air filters designed for this device. Do not operate without a filter. • There is no front or back orientation for the air filters.
NE-C801S-KDE_A_M_main.book Page 21 Friday, January 13, 2012 4:43 PM 7. Troubleshooting Check the following if your unit should fail during operation. You can also refer to the pages of this manual for complete instructions. Problem Cause Remedy Nothing happens when the power switch is pressed. Is the AC adapter connected Check that the plug is correctly to the an electrical inserted in an electrical outlet and the compressor? outlet. Unplug then reinsert the plug if necessary.
NE-C801S-KDE_A_M_main.book Page 22 Friday, January 13, 2012 4:43 PM 8. Technical data Name: COMP AIR Model: NE-C801KD (NE-C801S-KDE) Type: Compressor Nebuliser Rating (AC adapter): 100 - 240V ~ 350mA, 50/60Hz Rating (Compressor Nebuliser): 12V Operating Condition: Intermittent operation 20min. ON / 40min. OFF Operating Temperature/ Humidity: +10°C to +40°C 30% to 85% RH Storage Temperature/ Humidity/Air Pressure: -20°C to +60°C 10% to 95% RH 700 - 1060 hPa Weight: Approx.
NE-C801S-KDE_A_M_main.book Page 23 Friday, January 13, 2012 4:43 PM 8.Technical data Technical data for the OMRON NE-C801KD Compressor with the OMRON V.V.T. (Virtual Valve Technology) Nebuliser kit: Particle Size: **MMAD Approx. 3µm MMAD = Mass Median Aerodynamic Diameter Medication Tank Capacity: 7ml maximum Appropriate Medication Quantities: Sound: 2ml minimum - 7ml maximum Nebulisation Rate: *Approx. 0.3ml/min (by weight loss) Aerosol Output: **0.47ml (2ml, 1%NaF) Aerosol Output Rate: **0.
NE-C801S-KDE_A_M_main.book Page 24 Friday, January 13, 2012 4:43 PM 8.Technical data Important information regarding Electro Magnetic Compatibility (EMC) With the increased number of electronic devices such as PC’s and mobile (cellular) telephones, medical devices in use may be susceptible to electromagnetic interference from other devices. Electromagnetic interference may result in incorrect operation of the medical device and create a potentially unsafe situation.
NE-C801S-KDE_A_M_main.book Page 25 Friday, January 13, 2012 4:43 PM 9. Optional parts and accessories V.V.T. Nebuliser Kit N-01 & Mouthpiece Set V.V.T. Nebuliser Kit N-01 & Smile Bear Set V.V.T. Nebuliser Kit N-01 & Happy Bunny Set Model: C801-NSET3 Order No.: 9515573-6 Model: C801-NKKD1 Order No.: 9520011-1 Model: C801-NKKD2 Order No.: 9520012-0 Mouthpiece Child Mask (PVC) Infant Mask (PVC) Model: C28-3P Order No.: 9956273-5 Model: C28-9 Order No.: 9956276-0 Model: C801-MSSP Order No.
NE-C801S-KDE_A_M_main.book Page 26 Friday, January 13, 2012 4:43 PM Manufacturer OMRON HEALTHCARE Co., Ltd. 53, Kunotsubo, Terado-cho, Muko, Kyoto, 617-0002 JAPAN EU-representative OMRON HEALTHCARE EUROPE B.V. Scorpius 33, 2132 LR Hoofddorp, THE NETHERLANDS www.omron-healthcare.com Production facility OMRON (DALIAN) CO., LTD. Dalian, CHINA OMRON HEALTHCARE UK LTD. Opal Drive, Fox Milne, Milton Keynes, MK15 0DG, U.K. Subsidiary OMRON MEDIZINTECHNIK HANDELSGESELLSCHAFT mbH John-Deere-Str.