User Manual

51
12. Specifications
Blood pressure measurement
Cuff pressure range
0 to 299 mmHg
Blood pressure measurement range
SYS: 60 to 260 mmHg
DIA: 40 to 215 mmHg
Pulse measurement range
40 to 180 beats / min.
Accuracy
Pressure: ±3 mmHg
Pulse: ± 5 % of display reading
Inflation
Automatic by electric pump
Deflation
Automatic pressure release valve
Measurement method
Oscillometric method
EKG recording
Performance characteristics
EKG channel: Single channel
Input dynamic range: 10 mV Peak-to-Peak
Circuitry
Frequency response: 0.67 Hz to 40 Hz
CMRR: > 60 dB
Input Impedance: > 10 MOhm
Output
Modulation: Frequency modulated ultrasonic audio tone
Center frequency: 19 kHz
Modulation index: 200 Hz/mV