INSTRUCTION MANUAL ENGLISH ESPAÑOL 10 series Plus Blood Pressure Monitor with ComFitTM Cuff Model BP791IT
TABLE OF CONTENTS Before Using the Monitor Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4 Safety Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5 Operating the Device . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5 Risk of Electrical Shock . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7 Care and Maintenance . . . . . . . . . . . . . . . . .
TABLE OF CONTENTS Taking a Measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .30 Calibration Check System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .30 Cuff Wrap Guide . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .30 Using the Guest Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .31 Selecting the USER ID . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Before Using the Monitor INTRODUCTION Thank you for purchasing the Omron® BP791IT 10 series Plus Automatic Blood Pressure Monitor with ComFitTM Cuff. Fill in for future reference. DATE PURCHASED: __________________ SERIAL NUMBER: ____________________ • Staple your purchase receipt here • Register your product on-line at www.register-omron.com Your new blood pressure monitor uses the oscillometric method of blood pressure measurement.
SAFETY INFORMATION To assure the correct use of the product basic safety measures should always be followed including the warnings and cautions listed in this instruction manual. SAFETY SYMBOLS USED IN THIS INSTRUCTION MANUAL WARNING Indicates a potentially hazardous situation which, if not avoided, could result in death or serious injury.
SAFETY INFORMATION OPERATING THE DEVICE (continued) This device is intended for use in measuring blood pressure and pulse rate in the adult population. Do not use this device on infants or persons who cannot express their intentions. Read the Special Conditions section (page 37) of the instruction manual if your systolic pressure is known to be more than 220 mmHg. Inflating to a higher pressure than necessary may result in bruising where the cuff is applied. Operate the device only as intended.
SAFETY INFORMATION RISK OF ELECTRICAL SHOCK Do not plug or unplug the power cord into the electrical outlet with wet hands. Do not overload power outlets. Plug the device into the appropriate voltage outlet. CARE AND MAINTENANCE Do not subject the monitor to strong shocks, such as dropping the unit on the floor. Do not submerge the device or any of the components in water. Store the device and the components in a clean, safe location.
BEFORE TAKING A MEASUREMENT To ensure a reliable reading follow these recommendations: 1. Avoid eating, drinking alcohol, smoking, exercising, and bathing for 30 minutes prior to taking a measurement. Rest for at least 15 minutes prior to taking the measurement. 2. Stress raises blood pressure. Avoid taking measurements during stressful times. 3. Measurements should be taken in a quiet place. 4. Remove tight-fitting clothing from your arm. 5. Sit on a chair with your feet flat on the floor.
Operating Instructions KNOW YOUR UNIT Main Unit: Display Calibration Check System SET Button Morning/Evening Average ( ) Button USER ID Selection Switch Memory ( Up/Down ( ) Buttons / ) Button Start/Stop Button Battery Compartment USB Port Air Jack AC Adapter Jack 9
KNOW YOUR UNIT Components: Housing Air Tube Cord AC Adapter Air Plug (Connects to Air Jack) Arm Cuff 4 “AA” Alkaline Batteries Storage Case USB Cable INSTRUCTION MANUAL ENGLISH ESPAÑOL 10 series Plus Blood Pressure Monitor with ComFitTM Cuff Model BP791IT Instruction Manual Quick Start Guide 10
UNIT DISPLAY Morning Average Symbol Evening Average Symbol Date/Time Display Irregular Heartbeat Symbol USER ID Symbol (A or B) Average Value Symbol Week Display TruReadTM Symbol Memory Symbol Systolic Blood Pressure Morning Hypertension Symbol Diastolic Blood Pressure Blood Pressure Level Indicator Movement Symbol Deflation Symbol Heartbeat Symbol Low Battery Indicator Cuff Wrap Guide Pulse Display / Memory Number 11
DISPLAY SYMBOLS IRREGULAR HEARTBEAT SYMBOL ( ) When the monitor detects an irregular rhythm two or more times during the measurement, the Irregular Heartbeat Symbol ( ) will appear on the display with the measurement values. An irregular heartbeat rhythm is defined as a rhythm that varies by less than 25% from Pulse the average rhythm or 25% more than the Blood average rhythm detected while the monitor pressure is measuring the systolic and diastolic blood pressure.
DISPLAY SYMBOLS Internationally recognized guidelines for upper limit of normal home blood pressure readings are as follows: Home Guideline for Upper Limit of Normal BP Systolic Blood Pressure 135 mmHg Diastolic Blood Pressure 85 mmHg This is a general guideline because blood pressure depends on age and morbidity. A lower home blood pressure goal may be recommended for certain patients, including diabetic patients, pregnant women, and patients with renal failure.
DISPLAY SYMBOLS HEARTBEAT SYMBOL ( ) The Heartbeat symbol flashes on the display at every heartbeat during the measurement. AVERAGE VALUE SYMBOL ( ) The Average Value Symbol is displayed when you press the memory button. The most recent average reading appears on the display screen. MORNING AVERAGE SYMBOL ( ) The Morning Average symbol is displayed when viewing Morning Averages using the Memory Function.
USING THE AC ADAPTER IMPORTANT SAFETY INSTRUCTIONS SAVE THESE INSTRUCTIONS - DANGER TO REDUCE THE RISK OF FIRE OR ELECTRIC SHOCK, CAREFULLY FOLLOW THESE INSTRUCTIONS. For connection to a power supply not in the U.S.A., use an attachment plug adapter of the proper configuration for the power outlet. The power unit (AC Adapter) is intended to be correctly oriented in a vertical or floor mount position. NOTES: • Use only the authorized Omron AC Adapter that came with this monitor.
USING THE AC ADAPTER TO CONNECT THE AC ADAPTER ① Insert the AC Adapter Plug into the AC Adapter Jack on the back side of the monitor as illustrated. ② Plug the AC Adapter into a 120V AC outlet (60 cycles). Housing Cord TO DISCONNECT THE AC ADAPTER ① Hold the Housing to remove the AC Adapter from the electrical outlet. ② Remove the AC Adapter plug from the monitor. WARNING Do not plug or unplug the power cord into the electrical outlet with wet hands. WARNING Do not overload power outlets.
BATTERY INSTALLATION Battery Cover 1. Remove the battery cover by pulling the hook up while pushing it in the direction of the arrow. 2. Install 4 “AA” size batteries so the + (positive) and - (negative) polarities match the polarities of the battery compartment as indicated. 3. Replace the battery cover.
BATTERY INSTALLATION BATTERY REPLACEMENT Low Battery Indicator When the Low Battery Indicator appears on the display screen turn the monitor off and remove all the batteries. Replace with four new batteries at the same time. Longlife alkaline batteries are recommended. WARNING If battery fluid should get in your eyes, immediately rinse with plenty of clean water. Contact a physician immediately. NOTES: • Turn the unit off before replacing the batteries.
SETTING THE DATE AND TIME Set the monitor to the current date and time before taking a measurement for the first time. If the date and time are not set correctly, the measurement values stored in the memory and the average values may not be correct. If the date and time are not set, “-:--” appears during or after measurement. The Blood Pressure Monitor automatically stores up to 100 measurement values per user with the dates and times.
SETTING THE DATE AND TIME 1. TO START THE SETTING Press and hold the SET button to adjust the date and time. The year flashes on the display. 2. SETTING THE YEAR The year can be set between 2010 and 2040. When the display reaches 2040, it will return to 2010. Press the Up button ( by one year. ) to advance Press the SET button to set the current year. The month flashes on the display. NOTES: • Press the Down button ( values.
SETTING THE DATE AND TIME 3. SETTING THE MONTH Press the Up button ( by one month. ) to advance Press the SET button to set the current month. The day flashes on the display. 4. SETTING THE DAY Press the Up button ( by one day. ) to advance Press the SET button to set the current day. The hour flashes on the display.
SETTING THE DATE AND TIME 5. SETTING THE HOUR The time is set using AM or PM. Press the Up button ( by one hour. ) to advance Press the SET button to set the current hour. The minute flashes on the display. 6. SETTING THE MINUTE Press the Up button ( ) to advance by one minute. Press the SET button to set the minute setting. 7. TO TURN THE MONITOR OFF Press the START/STOP button.
SETTING TRUREADTM The TruReadTM Mode takes three consecutive measurements. The monitor will inflate, take a measurement, and deflate three times separated by a short interval of time between each measurement. 1. Select your USER ID (A or B). 2. Press the SET button. The TruReadTM symbol on the display. and the oFF setting appears 3. Press the Up or Down button ( / ) to select on. NOTE: Press the Up or Down button to alternate between on and oFF.
SETTING TRUREADTM 4. Press the SET button to confirm the change. If on is selected, the number of seconds between each measurement is displayed. NOTE: The default value between each measurement is 60 seconds. 5. Press the Up or Down button ( / time between each measurement. ) to change the interval of The interval can be set to 15 seconds, 30 seconds, 60 seconds or 120 seconds. Continue to press the Up or Down button until the desired interval appears on the display. 6.
SETTING TRUREADTM 7. If the date and time are correct, press the START/STOP button to turn the monitor off. To change the date and time, refer to “Setting the Date and Time”.
APPLYING THE ARM CUFF CAUTION Read all of the information in the instruction manual and any other literature in the box before operating the unit. CAUTION This device is intended for use in measuring blood pressure and pulse rate in the adult population. Do not use this device on infants or persons who cannot express their intentions. APPLYING THE CUFF ON THE LEFT ARM 1. Make sure the air plug is securely inserted in the main unit. 2. Remove tight-fitting clothing from your upper left arm.
APPLYING THE ARM CUFF 3. Sit on a chair with your feet flat on the floor. Place your left arm on a table so the cuff is level with your heart. Thumb Grip 4. Hold the thumb grip on the cuff securely with your right hand. Cloth Fastener 5. Turn the palm of your left hand upward.
APPLYING THE ARM CUFF 6. Apply the cuff to your left upper arm so the blue stripe is on the inside of your arm and aligned with your middle finger. The air tube runs down the inside of your arm. The bottom of the cuff should be approximately 1/2" above your elbow. Blue Stripe 1/2 inch (1 - 2 cm) Air Tube 7. Wrap the cuff firmly in place around your arm using the cloth fastener.
APPLYING THE ARM CUFF APPLYING THE CUFF ON THE RIGHT ARM When taking a measurement using the right arm use this instruction for Step 6 on page 28. Apply the cuff to your right upper arm so the thumb grip is centered on the inside of your inner arm. The bottom of the cuff should be approximately 1/2'' above your elbow. Thumb Grip NOTE: Be careful not to rest your arm on the air tube. This will restrict the flow of air to the cuff.
TAKING A MEASUREMENT The monitor is designed to take measurements and store the measurement values in the memory for two people using USER ID A and USER ID B. The monitor can also be used to take a single measurement for other persons using the GUEST Mode. The USER ID symbol is not displayed when using the GUEST Mode. Measurements taken using the GUEST Mode are not stored in the memory. WARNING Contact your physician for specific information about your blood pressure.
TAKING A MEASUREMENT USING THE GUEST MODE 1. Press and hold the START/STOP button. The A and B Symbols disappear from the display. A and B Symbols disappear 2. Release the START/STOP button. The cuff starts to inflate automatically. NOTE: To stop the inflation press the START/STOP button. The monitor will stop inflating, start deflation and turn off. 3. Inflation stops and the measurement is started. When the measurement is completed, your blood pressure and pulse rate appear on the display.
TAKING A MEASUREMENT SELECTING THE USER ID Select your USER ID. Always use the same USER ID when taking a measurement. The monitor stores the measurement values in the memory by the USER ID selected. These measurement values are used to calculate the Morning Averages and Evening Averages. Slide the USER ID Selection Switch right or left to select USER A or B.
TAKING A MEASUREMENT USING THE SINGLE MODE 1. Select your USER ID (A or B). 2. Press the START/STOP button. All symbols appear on the display. The USER ID Symbol you selected flashes on the display. The cuff starts to inflate automatically. As the cuff inflates, the monitor automatically determines your ideal inflation level. This monitor detects the pulse during inflation. Do not move your arm and remain still until the entire measurement process is completed.
TAKING A MEASUREMENT 3. Inflation stops automatically and the measurement is started. As the cuff deflates, decreasing numbers appear on the display. The Heartbeat Symbol ( ) flashes at every heartbeat. 4. When the measurement is complete, the arm cuff completely deflates. Your blood pressure and pulse rate are displayed. 5. Press the START/STOP button to turn the monitor off. NOTE: The monitor will automatically turn off after two minutes.
TAKING A MEASUREMENT USING THE TRUREADTM MODE 1. Select your USER ID (A or B). 2. Press the START/STOP button. All symbols appear on the display. The USER ID Symbol you selected flashes on the display. The cuff starts to inflate automatically. NOTE: To stop the inflation or measurement, press the START/STOP button. The monitor will stop inflating, start deflating and will turn off. 3. Inflation stops automatically and the measurement is started.
TAKING A MEASUREMENT The second measurement is complete, waiting to start the next measurement. The consecutive measurements are complete. The average for three measurements appears on the display. 5. Press the SET button to display the blood pressure and pulse rate for the first measurement. Press the SET button again to display the next measurement. 6. Press the START/STOP button to turn the monitor off. WARNING Contact your physician for specific information about your blood pressure.
TAKING A MEASUREMENT WARNING The monitor is not intended to be a diagnostic device. SPECIAL CONDITIONS CAUTION Inflating to a higher pressure than necessary may result in bruising where the cuff is applied. If your systolic pressure is known to be more than 220 mmHg, press and hold the START/STOP button until the monitor inflates 30 to 40 mmHg higher than your expected systolic pressure. NOTES: • Do not apply more pressure than necessary. • The monitor will not inflate above 299 mmHg.
USING THE MEMORY FUNCTION The monitor is designed to store the blood pressure and the pulse rate in the memory for two people (USER A and USER B) every time a measurement is completed. The monitor automatically stores up to 100 sets of measurement values (blood pressure and pulse rate) for each user (A and B). When 100 sets of measurement values are stored, the oldest record is deleted to save the most recent values.
USING THE MEMORY FUNCTION TO DISPLAY THE MEASUREMENT VALUES 1. Select your USER ID (A or B). 2. Press the Memory button ( The Average Symbol ( reading is shown. ).
USING THE MEMORY FUNCTION 3. Press the Down button ( ) to display the most recent set of measurement values on the screen. The values are displayed from the most recent to the oldest. NOTE: The date and time are alternately displayed with the measurement values. Alternating display Press the Up or Down button to display the next set of values. : To the older readings : To the more recent readings NOTE: Press and hold the Up or Down button to display the values faster.
USING THE MEMORY FUNCTION MORNING AND EVENING AVERAGES The monitor calculates and displays a weekly average for measurements taken in the morning ( ) and evening ( ). The monitor stores 8 weeks of Morning Averages and 8 weeks of Evening Averages for each user (A and B). NOTE: The week begins Sunday at 4:00 AM. MORNING AVERAGES Morning averages are based on the first average reading using the TruReadTM Mode or the first reading using the Single Mode. Measurement times must be between: 4:00 AM and 9:59 AM.
USING THE MEMORY FUNCTION TO DISPLAY MORNING AND EVENING AVERAGES 1. Select your USER ID (A or B). 2. Press the Morning/ Evening Average button ( ) to select the Morning or Evening average. The average for the current week “THIS WEEK” appears on the display. 3. Press the Down ( ) button to display the previous weeks. With each press of the Down ( ) button, the values are displayed from 1 week ago to the older. With each press of the Up ( ) button, the values are displayed from 7 weeks ago to the recent.
USING THE MEMORY FUNCTION DISPLAY COMBINATIONS In addition to the Morning Average and Evening Average Symbols, the monitor may also display the Morning Hypertension Symbol if the morning average for that week is above the internationally recognized guidelines. (See page 13 for details.) Depending on your measurement results, these may be displayed in the following combinations.
USING THE MEMORY FUNCTION The Morning Hypertension Symbol ( ) appears if the weekly average for morning measurements is above 135/85. In this case the Morning Hypertension Symbol ( ) is displayed when the Evening Average is displayed, regardless of the values for the Evening Average. TO DELETE ALL VALUES STORED IN THE MEMORY The values stored in the memory are deleted by USER ID. You cannot partially delete values stored in the memory. All values for the user you select will be deleted. 1.
Downloading Instructions INSTALLING THE SOFTWARE Optional Instructions for PC Connection To install software, choose either Microsoft® HealthVaultTM or Omron Health Management software. NOTE: Any measurement values taken before setting the date and time are not transferred to Microsoft HealthVaultTM and Omron Health Management software. DOWNLOADING AND INSTALLING MICROSOFT® HEALTHVAULTTM HealthVault is a free Web-based platform designed to put people in control of their health data.
INSTALLING THE SOFTWARE OPERATING ENVIRONMENT REQUIREMENTS The following operating environments are required to run Microsoft® HealthVault.
INSTALLING THE SOFTWARE DOWNLOADING AND INSTALLING OMRON HEALTH MANAGEMENT SOFTWARE Omron Health Management Software allows you to view, process and print data about blood pressure and walking measured by Omron’s compatible blood pressure monitor and pedometer. • To install Omron Health Management Software onto your PC, you must have administrative privileges. • Close all applications currently running before you install the software.
INSTALLING THE SOFTWARE OPERATING ENVIRONMENT REQUIREMENTS The following operating environments are required to run the Omron Health Management software.
USING THE SOFTWARE For help using the software, double-click [Omron Health Management Software Users Manual.pdf] shortcut on desktop, or click [Help] in this software after starting up. 1. Double-click the [Omron Health Management Software] shortcut icon on the desktop. If no icon appears on the desktop: From the [Start] menu, select [All programs] or [Programs] [Omron Health Management Software], and click the [Omron Health Management Software] shortcut. 2.
USING THE SOFTWARE 3. In the New User Registration box, enter your name (e.g., Chuck) and click [OK]. NOTES: • Omron Health Management Software can manage multiple users’ data. Data is controlled by user name. • The number of users who can register for this software varies depending on the amount of free disc space in your PC. • The user name can be up to 28 characters. 4. Click the user name, and click [Graph Screen]. 5. Click the [Blood Pressure Monitor] icon.
USING THE SOFTWARE Use only the authorized Omron USB cable, Model HHX-CABLEUSB2, included with this unit. CAUTION Use only Omron authorized parts and accessories. Parts and accessories not approved for use with the device may damage the unit. 1. Open the USB Port cover. 2. Plug the normal USB Connector (Large Connector) into the personal computer. Normal USB Connector Mini USB Connector 3. Plug the mini USB connector (Small Connector) into the USB Port.
USING THE SOFTWARE 4. Click [Download Data] icon to import the data. 5. The Confirm dialog appears. Make sure your cable is connected then click [Yes]. 6. Select the blood pressure monitor being used and click [OK]. 7. Start importing data. Click [Cancel] if you want to cancel. 8. The dialog appears when complete. Click [OK] and view your data on the graph.
Care and Maintenance CARE AND MAINTENANCE To keep your blood pressure monitor in the best condition and protect the unit from damage, follow the directions listed below: Keep the monitor in the storage case when not in use. Make sure the AC adapter is placed under the main unit so that it does not damage the display. Avoid kinking or sharply bending the AC Adapter cord. Do not forcefully bend the arm cuff or air tube. Do not fold tightly. Clean the monitor with a soft dry cloth.
CARE AND MAINTENANCE Use the unit consistent with the instruction provided in this manual. CAUTION Changes or modification not approved by Omron Healthcare will void the user warranty. Do not disassemble or attempt to repair the unit or components. CAUTION Use only Omron authorized parts and accessories. Parts and accessories not approved for use with the device may damage the unit.
ERROR INDICATORS AND TROUBLESHOOTING TIPS ERROR INDICATORS SYMBOL CAUSE CORRECTION Air plug disconnected. Insert the plug securely. Refer to “Applying the Arm Cuff”. Arm cuff not applied correctly. Apply the arm cuff correctly. Refer to “Applying the Arm Cuff”. Air is leaking from the arm cuff. Replace the cuff with the new one. Repeat measurement. Remain still and do not talk during measurement. Refer to “Taking a Measurement”. Cuff over or under inflated.
ERROR INDICATORS AND TROUBLESHOOTING TIPS ERROR INDICATORS SYMBOL CAUSE CORRECTION Replace the four batteries. Refer to “Battery Installation”. Batteries are worn. TROUBLESHOOTING TIPS PROBLEM CAUSES AND SOLUTIONS No power. Replace all four batteries with new ones. No display appears on the unit. Check the battery installation for proper placement of the battery polarities. Measurement values appear too high or too low. Blood pressure varies constantly.
FCC STATEMENT NOTE: POTENTIAL FOR RADIO/TELEVISION INTERFERENCE (for U.S.A. only) This product has been tested and found to comply with the limits for a Class B digital device, pursuant to part 15 of the FCC rules. These limits are designed to provide reasonable protection against harmful interference in a residential installation.
LIMITED WARRANTY Your Omron® BP791IT 10 series Plus Automatic Blood Pressure Monitor, excluding the arm cuff and batteries, is warranted to be free from defects in materials and workmanship appearing within 5 years from the date of purchase, when used in accordance with the instructions provided with the monitor.
SPECIFICATIONS Model: Display: Measurement Range: Accuracy/Calibration: BP791IT (HEM-7222-ITZ) LCD Digital Display Pressure: 0 to 299 mmHg, Pulse: 40 to 180/min. Pressure: ±3mmHg or 2% of reading Pulse: ±5% of reading Inflation: Automatic by electric pump Deflation: Automatic pressure release valve Rapid Air Release: Automatic exhaust valve Measurement Method: Oscillometric method Power Source: 120V 60Hz AC adapter or 4 “AA” batteries Battery Life: Approx.
Manufactured by OMRON HEALTHCARE Co., Ltd. Made in China Distributed by: Omron Healthcare, Inc. 1200 Lakeside Drive Bannockburn, IL 60015 www.omronhealthcare.com Copyright © 2011 Omron Healthcare, Inc.