User Manual

191
Platinum Resistance Thermometer Calibration (Thermocouple/Resistance Thermometer Input)
Section 6-4
7. Next calibrate the B-B' input.
Change the connections as follows:
8. When the M key is pressed, the status changes as shown to the left.
Set the 6-dial to 10
.
Allow the count value on the No. 2 display to fully stabilize, then press the
D key to temporarily register the calibration settings.
If this count value is outside of the specified range, the No. 2 display will
flash and the count value will not be temporarily registered.
9. When the M key is pressed, the status changes as shown to the left.
Short-circuit the 6-dial terminals to set 0
.
Note The 6-dial terminals must be short-circuited, because it is otherwise
impossible to set 0
for the 6-dial.
Allow the count value on the No. 2 display to fully stabilize, then press the
D key to temporarily register the calibration settings.
If this count value is outside of the specified range, the No. 2 display will
flash and the count value will not be temporarily registered.
10. When the M key is pressed, the status changes as shown to the left.
The data to be temporarily registered is not displayed if it is not complete.
Press the U key. The No. 2 display changes to yes. Release the key and
wait two seconds or press the M key. This stores the temporarily regis-
tered calibration data to EEPROM.
To cancel the saving of temporarily registered calibration data to EE-
PROM, press the M key (while no is displayed in the No. 2 display) with-
out pressing the U key.
11. The calibration mode is quit by turning the power OFF.
E5CN
4
3
A
B'
B
5
10
9
E5CN-U
2
3
A
B'
B
1
11
10
6-dial
6-dial
Input power supply
Input power supply
E5AN/EN
17
16
A
B'
B
18
2
1
6-dial
Input power supply
pb10
4544
pb-0
4083
str
no