Owner's manual

General Testing:
1. Release the clamps on front door to have access to the test section.
2. Mount your specimen (e.g., component, PCB or heat sink) in the test section.
3. Clamp the door back on the WT-3106.
4. Place flow measurement instrument in the instrument port.
5. Connect the fan cable to the controller box.
6. Connect the fan controller box to a DC power supply or wind tunnel controller.
Important Note: Voltage to the fans cannot exceed 24 Volts-DC.
7. Adjust voltage until desired flow rate is attained.
Component and Heat Sink Testing:
1. Release the clamps on front door to have access to the test section.
2. Mount your component(s) in the test section.
3. Clamp the door back on the WT-3106.
4. Connect the fans to a variable DC power supply. Voltage to the fans cannot exceed 24
Volts-DC. Adjust voltage until desired flow level is attained.
5. Insert the velocity and temperature probes into the instrument ports provisioned in
upstream of the test section.
6. Turn on the power supplies and start the experiment.
SYSTEM OPERATION