Owner's manual

%
2) Using a millivolt calibrator or the transducer output itself,
set the gain so that the proper full scale output voltage is
realized (the mV calibrator or transducer should be set to
simulate full scale output).
3) If system offsets must be accounted for repeat step 1
again with the inputs disconnected from the source and
connected to ground, or short them together with the
bridge connected, then reconnect the inputs and re-zero
the output with the bridge balance (if used).
4) Steps 1 - 3 above may need to be repeated several times
to achieve the desired accuracy of gain and offset.
FIGURE 6. Full Bridge with No Remote Sense
FIGURE 4. OM2-163 Trimpot Adjustment Detail
FIGURE 5. Full Bridge with Remote Excitation Sense
OM2-163 Application Examples