Owner's manual
OM-USB-TEMP User's Guide Specifications
26
Current excitation outputs (Ix+)
Table 16. Current excitation output specifications
Parameter Conditions Specification
Configuration 4 dedicated pairs:
±I1 - CH0/CH1
±I2 - CH2/CH3
±I3 - CH4/CH5
±I4 - CH6/CH7
Thermistor 10 µA typ.Current excitation output ranges
RTD 210 µA typ.
Tolerance ±5% typ.
Drift 200 ppm/°C
Line regulation 2.1 ppm/V max.
Load regulation 0.3 ppm/V typ.
Output compliance voltage
(relative to GND pins 9, 19, 28, 38)
3.90 V max.
-0.03 V min.
Note 18: The OM-USB-TEMP has four current excitation outputs, with ±I1 dedicated to the CH0/CH1
analog inputs, ±I2 dedicated to CH2/CH3, ±I3 dedicated to CH4/CH5, and ±I4 dedicated to
CH6/CH7. The excitation output currents should always be used in this dedicated configuration.
Note 19: The current excitation outputs are automatically configured based on the sensor (thermistor or
RTD) selected.
Environmental
Table 17. Environmental specifications
Operating temperature range 0 to 70 ° C
Storage temperature range -40 to 85 ° C
Humidity 0 to 90% non-condensing
Mechanical
Table 18. Mechanical specifications
Dimensions 127 mm (L) x 88.9 mm (W) x 35.56 (H)
User connection length 3 meters max.